Daftar Isi /
Table of Content
|
Par. |
Uraian |
Par. |
Description
|
Page
|
1 |
Tujuan |
1 |
Purpose
|
3
|
2 |
Ruang Lingkup |
2 |
Scope
|
3
|
3 |
Laporan |
3 |
Records
|
3
|
4 |
Referensi |
4 |
References
|
4
|
5 |
Definisi – Definisi |
5 |
Definition |
4 |
6 |
Qualifikasi dan Tanggung Jawab Personnel
|
6 |
Personnel Qualification and Responsibility
|
5
|
7 |
Perlengkapan
|
7 |
Equipment
(s)
|
5
|
7.1 |
Alat – alat
Ultrasonic
|
7.1 |
Ultrasonic
Equipments
|
5
|
7.2 |
Search
units
|
7.2 |
Search
units
|
5
|
7.3 |
Couplant
|
7.3 |
Couplant
|
6
|
7.4 |
Blok
Kalibrasi
|
7.4 |
Calibration
Block
|
7
|
8 |
Persyaratan
Keselamatan
|
8 |
Safety
Requirement
|
8
|
9 |
Instruksi – Instruksi |
9 |
Instructions |
8
|
9.1 |
Persiapan permukaan
|
9.1 |
Surface Preparation.
|
8 |
9.2
|
Sistem Kalibrasi
|
9.2 |
Calibration Systems
|
8
|
9.2.1
|
Persyaratan Umum
|
9.2.1 |
General Requirement
|
8
|
9.2.2
|
Teknik Kalibrasi probe sudut
|
9.2.2 |
Angel Beam Calibration technique
|
9
|
9.2.3
|
Kalibrasi probe normal
|
9.2.3 |
Straight beam calibration.
|
10
|
9.2.4
|
Konfirmasi Kalibrasi
|
9.2.4 |
Calibration
Confirmation
|
11
|
9.2.5
|
Pengujian
|
9.2.5 |
Examination
|
11
|
9.2.6
|
Evaluasi ukuran cacat.
|
9.2.6 |
Discontinuity size evaluation
|
13
|
9.2.7
|
Acceptance standard
|
9.2.7 |
Acceptance standards
|
13
|
9.2.8
|
Pengujian hasil perbaikan
|
9.2.8 |
Examination of repairs
|
14
|
1 |
Tujuan |
1 |
Purpose
|
|||||||||
a. |
Prosedur ini untuk digunakan sebagai petunjuk pelaksanaan Tes Ultrasonic Kontak Langsung. |
a. |
This
procedure provides guidance during the conduct of Direct Contact Ultrasonic
Test.
|
|||||||||
b. |
Prosedur ini terdiri dari versi Inggris dan Indonesia, jika ada konflik antara keduanya maka versi Inggris yang dipakai |
b. |
This procedure consists of English and
Indonesian versions, should there is conflict between these versions the
English version shall govern.
|
|||||||||
|
|
|
|
|||||||||
2 |
Ruang Lingkup |
2 |
Scope
|
|||||||||
2.1 |
Prosedur ini menyajikan syarat 2 dan
standar acceptance yang digunakan untuk pengujian dengan Ultrasonic kontak
pada:
|
2.1 |
This
procedure provides the requirements and acceptance standards for the contact
ultrasonic examination for
|
|||||||||
a. |
lasan full penetrasi ferrous material (carbon steel, low alloy steel
dan stainless steel material) dengan ketebalan antara 0.5 in. (13.0 mm)
sampai 10 in. (250 mm).
Lasan fillet yang tanpa ada groove nya tidak termasuk
dalam procedure ini.
Sambungan las mungkin terbuat dari pipa, tube atau
plat.
|
a. |
the full penetration welds of ferrous material
(carbon steel, low alloy steel and stainless steel material) with thickness
between 0.5 in. (13.0 mm) upto 10 in. (250 mm).
The fillet weld without groove is not included in this
procedure
Weld
joints may be made from pipes, tubes
or plates.
|
|||||||||
b. |
base material ferrous material (carbon steel, low alloy
steel dan stainless steel material) dengan ketebalan antara 0.5 in. (13.0 mm)
sampai 10 in. (250 mm) untuk pengujian
laminasi atau defect lainnya.
|
b. |
the base materials of ferrous material (carbon steel, low alloy steel
and stainless steel material) with thickness between 0.5 in. (13.0 mm) upto 10 in. (250 mm) for
examination the lamination or other defects.
|
|||||||||
2.2 |
Qualifikasi Procedure harus dilaksanakan jika ada perubahan essential variable dari persyaratan di Table T – 421; jika perubahan nonessential variable dari persyaratan di Table T – 421 tidak diperlukan requalifikasi procedure. Semua perubahan dari essential atau non essential variable pada nilai, range nilai, yang disebut dalam prosedur harus diperlukan revisi pada, atau dengan addendum untuk prosedur yang tertulis. |
2.2 |
Procedure
Qualification shall be made when a change essential variable of a requirement
in Table T - 421; a change
nonessential variable of a requirement in Table T – 421 need not requalification of written procedure. All
changes of essential or non essential variable from the value, or range
value, specified by the written procedure shall require revision of, or an
addendum to, the written procedure.
|
|||||||||
|
|
|
|
|||||||||
3 |
Laporan dan Catatan |
3 |
Records
|
|||||||||
3.1 |
Hasil dari Pengujian Ultrasonic harus di report dalam
Form Ultrasonic Test Report yang
tersedia (terlampir).
|
3.1 |
Ultrasonic Examination Results shall be reported on
the provided Ultrasonic Test Report Form
(attached)
|
|||||||||
3.2 |
Semua pantulan dari area yang tidak
dibetulkan, mempunyai respon melebihi 50 % dari reference level harus direcord dalam Form Ultrasonic
Test Report, record ini harus
menyebutkan setiap area, respond level, dimensi, dalamnya dari permukaan dan
klasifikasi.
|
3.2 |
All reflections from uncorrected areas
having responses that exceed 50% of the reference level shall be recorded on
the attached record form, this record
shall indicate each area, the response level, the dimension, the depth bellow
the surface, and the classification.
|
|||||||||
3.3 |
Ultrasonic Reports harus disimpan selama minimum 5 (lima) tahun. |
3.3 |
Ultrasonic Reports shall be filed and
maintained minimum 5 (five) years.
|
|||||||||
|
|
|
|
|||||||||
4 |
Referensi |
4 |
References
|
|||||||||
4.1 |
Written
Practice Procedure No.:
SMS – 601 – P – 001.
|
4.1 |
Written
Practice Procedure No.: SMS – 601 – P – 001.
|
|||||||||
4.2 |
ASME
2007 Code Section I, V article 4, VIII Div. 1, VIII Div.
2 dan ASME B31.1
|
4.2 |
ASME
2007 Code Section I, V article 4, VIII Div. 1, VIII Div.
2 and ASME B31.1
|
|||||||||
4.3 |
API 650; Welded
Steel Tanks for Oil Storage.
|
4.3 |
API 650; Welded
Steel Tanks for Oil Storage.
|
|||||||||
|
|
|
|
|||||||||
5 |
Definisi – Definisi |
5 |
Definitions |
|||||||||
5.1
|
Primary
reference response (level): respond dari ultrasonic dari refector di blok kalibrasi
pada jarak sound path tertentu, secara electronic diatur sampai persen
tertentu dari ketinggian screen.
|
5.1
|
Primary
reference response (level): the ultrasonic response from the basic calibration
reflector at the specified sound path distance, electronically adjusted to a
specified percentage of the full screen height.
|
|||||||||
5.2
|
D A C: kurva koreksi jarak dan
amplitudo.
|
5.2
|
D A C: Distance Amplitude
Correction curve
|
|||||||||
5.3
|
Probe
(Probe):
sebuah alat electro acoustic digunakan untuk meneruskan atau menerima
ultrasonic energy.
|
5.3
|
Search Unit
(Probe):
an Electro acoustic device used to transmit or receive ultrasonic energy
|
|||||||||
5.4
|
Transducer: sebuah alat electro
acoustical untuk merubah energi listrik ke energi getaran dan sebaliknya.
Probe or
Transducer
akan digunakan didalam prosedur ini dengan pengertian yang sama.
|
5.4
|
Transducer: an electro acoustical
device for converting electrical energy into acoustical energy and vice
versa.
Probe or
Transducer will
be used in this procedure with the
same meaning.
|
|||||||||
5.5
|
Kalibrasi: hubungan dari respond
sistem ultrasonic dengan kalibrator.
|
5.5
|
Calibration: correlation of the
ultrasonic system respond (s) with calibration reflector (s)
|
|||||||||
5.6
|
Straight
beam:
sebuah gelombang getaran pulsa yang berjalan normal (tegak lurus) dengan
permukaan yang di tes.
|
5.6
|
Straight
beam: a
vibrating pulse wave train travelling normal to the test surface.
|
|||||||||
5.7
|
Angle beam: sebuah terminologi yang
digunakan untuk menyatakan sebuah sudut datang atau pembengkokan selain tegak
lurus dengan permukaan yang di tes, hanya beam sudut gelombang shear yang
digunakan dalam prosedur ini.
|
5.7
|
Angle beam: a term used to describe
an angle of incidence or refraction other than normal to the surface of the
test object, only angle beam shear waves shall be used in this procedure.
|
|||||||||
|
|
|
|
|||||||||
|
|
|
|
|||||||||
|
|
|
|
|||||||||
6
|
Kualifikasi
dan Tanggung Jawab Personnel
|
6
|
Personnel
Qualification and Responsibility
|
|||||||||
6.1
|
Personel yang melaksanakan pengujian sesuai dengan
requiremen prosedure ini harus diqualifikasi dan disertifikasi
|
6.1
|
Personnel performing examinations to the requirements
this procedure shall be qualified and certified
|
|||||||||
6.2
|
Personel yang melaksanakan tes dan mengevaluasi
dengan prosedure ini harus memiliki Level II atau Level III.
|
6.2
|
Personnel performing examinations and evaluation to
the requirements this procedure shall be Level II or Level III
|
|||||||||
|
|
|
|
|||||||||
7
|
Perlengkapan
|
7
|
Equipments
|
|||||||||
7.1
|
Alat – alat
Ultrasonic
|
7.1
|
Ultrasonic
Equipments
|
|||||||||
7.1.1
|
Frekwensi
|
7.1.1
|
Frequency
|
|||||||||
|
Pengujian ini harus dilaksanakan dengan menggunakan
instrumen ultrasonic pulsa gema yang bisa menimbulkan frekwensi minimum
antara 1 MHz sampai 5 MHz dan harus mempunyai gain kontrol step 2.0 dB atau
kurang.
Instrumen dengan frequensi lain bisa digunakan, jika
dapat didemonstrasikan sensitivitasnya setara atau lebih baik dan
didokumentasikan.
|
|
This Examination shall be conducted with a pulse-echo
ultrasonic instrument capable of generating frequencies over the range of at
least 1 MHz to 5 MHz and shall be equipped with a stepped gain control in
units of 2.0 dB or less.
Instrument operating at other frequencies may be used,
if equal or better sensitivity can be demonstrated and documented
|
|||||||||
7.1.2
|
Lineariti
ketinggian Screen
|
7.1.2
|
Screen height
linearity
|
|||||||||
|
Ultrasonic instrumen harus menunjukkan linearitas
ketinggian screen sesuai dengan ASME Section V.
|
|
The ultrasonic instrument shall provide screen height
linearity in accordance with ASME Section V.
|
|||||||||
7.1.3
|
Lineariti control
Amplitudo
|
7.1.3
|
Amplitude control
linearity
|
|||||||||
|
Instrumen ultrasonic harus menggunakan control
akurasi amplitudo dalam range
± 20% dari perbandingan nominal amplitudo, untuk memungkinkan mengukur
indikasi2 diluar range dari vertikal display pada screen. Prosedur
yang mengevaluasi lineariti control amplitudo sesuai dengan ASME Section V.
|
|
The ultrasonic instrument shall utilize an amplitude
control accurate over it’s useful range to ± 20% of the nominal amplitude
ratio, to allow measurement of indications beyond the linear range of the
vertical display on the screen. The procedure for evaluating amplitude
control linearity in accordance with ASME Section V.
|
|||||||||
|
|
|
|
|||||||||
7.2
|
Search units
|
7.2 |
Search units
|
|||||||||
7.2.1
|
Frequensi dari transducer harus diantara 1 MHz sampai 5
MHz.
|
7.2.1 |
The frequencies of the transducer shall be within 1 Mhz
and 5 Mhz.
|
|||||||||
7.2.2
|
Pengujian harus
dilaksanakan dengan mode gelombang shear dari probes yang sudut nominalnya di
material carbon steel and low alloy steel adalah 45, 60, 70 derajat dan
mode gelombang longitudinal probe normal dengan teknik kontak langsung.
Ukuran probe yang
digunakan, probe sudut 8 x 9 mm atau 20 x 22 mm dan probe normal dengan
diameter 10 sampai 25 mm.
Probe dengan ukuran yang selain yang di sebut diatas
bisa digunakan, jika dapat didemonstrasikan sensitivitasnya setara atau lebih
baik dan didokumentasikan.
|
7.2.2 |
Examination shall be
performed with shear wave mode angle beam search units whose nominal angles
in the carbon steel and low alloy steel material are 45, 60, 70 deg.
and longitudinal wave mode straight beam search unit using direct contact
technique.
Angle beam search units dimension 8
x 9 mm or 20 x 22 mm and straight beam search diameter 10 upto 25 mm
shall be used.
The other stated search
unit dimensions may be used, if equal or better sensitivity can be
demonstrated and documented
|
|||||||||
7.2.3
|
Jika
penghitungan untuk evaluasi, maka sudut untuk stainless steel seperti dibawah
ini:
|
7.2.3 |
When
calculating for evaluation, the angle for stainless steel shall be as
follows:
|
|||||||||
|
Sudut di
Carbon and Alloy st.
70°
60°
45°
|
Sudut di
Stainless steel
65 °
57 °
43 °
|
|
Angle in
Carbon and Alloy st.
70°
60°
45°
|
Angle in
Stainless steel
65 °
57 °
43 °
|
|||||||
7.2.4
|
Direkomendasikan
untuk menambah sepatu (sepatu 3 mm mika) pada probe, dengan menggunakan lem
Super glue, sebelum mulai proses penyetelan. Kemudian yang digosok adalah
sepatu flexi glasnya.
Harus
hati – hati supaya tidak ada udara yang terjebak ketika proses pengeleman
tersebut.
|
7.2.4 |
It is recommended
that an extra wear plate (3 mm plexi glass probe shoe) be attached to
the probe, using Eastman / Super glue, before starting the fitting process.
Fitting then takes place using the Plexiglas probe shoe.
Care is to be
taken that no bubbles are trapped in the glue where the shoe adheres to the
probe.
|
|||||||||
7.2.5
|
Toleransi
yang dizinkan dari sudut beam adalah +2° / -0°. Jika deviasi dari sudut beam
melebihi toleransi ini, harus dibetulkan dengan cara menggerinda salah satu
ujung dari permukaan sepatu
|
7.2.5 |
A
tolerance of +2° / -0° in beam angle is permissible. Where the beam angle
deviation exceeds this tolerance, it is to be corrected by grinding more at
one end of the wear probe shoe.
|
|||||||||
7.2.6
|
Pemilihan probe
|
7.2.6 |
Search Unit selection
|
|||||||||
|
Tebal material
|
Sudut Probe
|
|
Material Thickness
|
Search Unit
Angle
|
|||||||
|
£ 30 mm
|
60 0 dan/atau 70 0
|
|
£ 30 mm
|
60
0 and / or 70 0
|
|||||||
|
> 30 mm
|
45
0 dan 60 0 atau 45 0
dan 70 0
|
|
> 30 mm
|
45 0 and 60 0 or
45 0 and 70 0
|
|||||||
|
|
|
|
|
|
|||||||
7.2.7
|
Penggunaan
kombinasi probe selain yang disebut diatas
diizinkan, asalkan seluruh volume dari part yang di uji bisa tercakup.
|
7.2.7 |
Using
other search unit combinations as mentioned above are permitted, provided the
whole volume of part examined can be covered
|
|||||||||
|
|
|
|
|||||||||
7.3
|
Couplant
|
7.3 |
Couplant
|
|||||||||
7.3.1
|
Couplant
yang mempunyai karakteristik pelicin seperti oli mesin # 20 to # 30, grease
atau C.M.C. harus digunakan.
Couplant
yang dipakai untuk pengujian harus sama dengan couplant yang dipakai
pada kalibrasi.
|
7.3.1 |
Couplant
having good wetting characteristics such as # 20 to # 30 machine oil, grease
or C.M.C. shall be used . The couplant
for examination shall be the same with the couplant used in
calibration.
|
|||||||||
7.3.2
|
Setelah
pengujian Couplant harus dibersihkan dengan kain atau tisu.
|
7.3.2 |
Couplant
should be removed by wiping with a cloth or absorbent paper after
examination.
|
|||||||||
|
|
|
|
|||||||||
7.4 |
Blok
Kalibrasi
|
7.4 |
Calibration
Block
|
|||||||||
7.4.1 |
Standar Blok kalibrasi
|
7.4.1 |
Basic Calibration Block
|
|||||||||
|
Blok
kalibrasi reflektor harus digunakan untuk menentukan primary reference response dari alat. Blok kalibrasi reflektor
bisa di tempatkan di salah satu komponen atau di standar blok kalibrasi.
Kalau tebal blok ±
25,4 mm berada antara dua range ketebalan seperti diperlihatkan di gambar 1,
blok yang digunakan harus bias diterima di porsi tersebut dari setiap
ketebalan yang dicover oleh 25,4 mm.
|
|
The
basic calibration reflectors shall be used to establish a primary reference
response of the equipment. The basic calibration reflectors may be located
either in the component material or in a basic calibration block. Where the
block thickness ±
25,4 mm spans two of the weld thickness range shown in Fig. 1, the block used
shall be acceptable in those portion of each thickness range covered by 25,4
mm.
|
|||||||||
7.4.2
|
Blok Material
|
7.4.2 |
Block Material
|
|||||||||
a.
|
Material
dari blok kalibrasi harus dibuat dari bentuk produk yang sama, dan
spesifikasi material yang sama atau P – number yang setara dengan material
yang akan di uji (dites). Yang dimaksud di paragraf ini, P no. 1, 3, 4 dan 5
dianggap setara.
|
a. |
The
material from which the block is fabricated shall be of the same product
form, and material specification or equivalent P-number grouping of the
materials being examined. For the purposes of the paragraph, P-Nos. 1.3.4.
and 5 materials are consider equivalent.
|
|||||||||
b.
|
Kalibrasi
blok untuk material yang berbeda, pemilihan material harus berdasarkan pada
material yang dimana pengujian akan dilaksanakan; jika pengujian akan
dilaksanakan dari kedua belah sisi reflektor kalibrasi harus dibuat dengan
kedua jenis material tersebut. Jika dua atau lebih ketebalan dalam suatu
pengujian, reflektor kalibrasi harus ditentukan dengan rata – rata dari
ketebalan lasan tsb.
|
b. |
For
calibration blocks for dissimilar metals weld, the material selection shall
be base on the material on the side of the weld from which the examination
will be conducted; if the examination will be conducted from both side,
calibration reflector shall be provided in both materials. Where two or more
base material thickness are involve, the calibration reflector shall be
determined by the average thickness of the weld.
|
|||||||||
c.
|
Blok
kalibrasi harus di temper minimum sesuai dengan yang dibutuhkan dari spesifikasi
material untuk type dan grade material tersebut, dan harus di post weld heat
treatment minimum 2 jam, jika blok kalibrasi mengandung lasan.
|
c. |
The
calibration block shall receive at least the minimum tempering treatment
required by the material specification for the type and grade, and also a
post weld heat treatment of at least 2 hours, if the calibration block
contains weld (s).
|
|||||||||
d.
|
Permukaan
dari blok kalibrasi harus mewakili dari permukaan dari komponen yang diuji.
|
d. |
The
finish on the surfaces of the block shall be representative of the surface
finish on the components.
|
|||||||||
e.
|
Blok
kalibrasi harus di uji dengan probe normal. Area yang mengandung indikasi
cacat melebihi refleksi dari backwall harus di buang dari jarak beam yang
dibutuhkan, untuk menghindari reflektor yang bervariasi dan membingungkan.
|
e. |
The
calibration block material shall be completely examined with a straight beam
search unit. Area that contains an indication exceeding the remaining back
reflection shall be excluded from the beam paths required to reach the
various calibration reflectors.
|
|||||||||
7.4.3
|
Lasan yang memanjang di material
dengan diameter lebih besar dari 20 inci (508 mm).
|
7.4.3 |
Longitudinal Weld in materials with
diameters greater than 20 inches (508 mm).
|
|||||||||
|
Untuk
pengujian pada material dimana surface diameter lebih besar dari 508 mm,
harus menggunakan blok yang mendekati bentuk curva-nya, atau alternatif
dengan flat (lihat Fig. 1).
|
|
For
examination in materials where the examination surface diameter is greater
than 508 mm, a block of essentially the same curvature, or alternatively,
flat basic calibration block (see Fig. 1), shall be used.
|
|||||||||
7.4.4
|
Lasan yang memanjang di material
dengan diameter 20 inci (508 mm) dan lebih kecil.
|
7.4.4 |
Longitudinal weld in materials with
diameters 20 inches (508 mm) and less.
|
|||||||||
|
Blok
kalibrasi harus dibentuk curva untuk material dengan diameter 20 inches (508
mm) dan lebih kecil. Kecuali yang di nyatakan dalam paragrap ini, single blok
kalibrasi bentuk curva boleh digunakan untuk mengkalibrasi surface benda uji
dengan range dari curva mulai 0.9 sampai 1.5 kali blok kalibrasi diameter.
|
|
The
basic calibration block shall be curved for materials with diameters 508 mm
and less. Except where otherwise states in this article, a single curved
basic calibration block may be used to calibrate the examination on surfaces
in the range of curvature from 0.9 to 1.5 times the basic calibration
block diameter.
|
|||||||||
|
|
|
|
|||||||||
8
|
Persyaratan Keselamatan
|
8 |
Safety Requirements
|
|||||||||
|
Peraturan
Safety
harus ditaati selama dalam pengujian.
|
|
Safety Regulation shall be followed
during examination.
|
|||||||||
|
|
|
|
|||||||||
9 |
Instruksi – instruksi
|
9 |
Instructions
|
|||||||||
9.1 |
Persiapan permukaan
|
9.1 |
Surface Preparation.
|
|||||||||
9.1.1
|
Base metal (permukaan kontak)
|
9.1.1 |
Base metal (contact surface)
|
|||||||||
|
Base metal pada setiap
sisi lasan harus bebas dari spater, permukaan yang tidak teratur, atau benda
asing yang mungkin mengganggu dalam pengujian.
|
|
The base metal on each
side of the weld shall be free of weld spatter, surface irregularities, or
foreign matter that might interfere with the examination.
|
|||||||||
9.1.2
|
Weld metals.
|
9.1.2 |
Weld metals.
|
|||||||||
|
Jika permukaan lasan mengganggu dalam pengujian,
lasan harus di persiapkan secukupnya untuk tidak mengganggu dalam pengujian
(sesuai peraturan permukaan bisa di buat / digerinda sampai rata dengan
permukaan base metal).
|
|
Where the weld surfaces interfere with the
examination, the weld shall be prepared as needed to permit examination (as a
rule, weld surface should be finished up to the same surface as the base
metal).
|
|||||||||
|
|
|
|
|||||||||
9.2
|
Sistem Kalibrasi
|
9.2 |
Calibration
Systems
|
|||||||||
9.2.1
|
Persyaratan Umum
|
9.2.1 |
General Requirement
|
|||||||||
a.
|
Kalibrasi
harus meliputi semua sistem pengujian dari alat ultrasonic. Kalibrasi yang
asli harus dilaksanakan pada kalibrasi blok. Pengecekan harus dipakai untuk
verifikasi kalibrasi jarak range dan jarak - amplitudo koreksi, pengecekan
harus meliputi semua sistem pengujian dari alat ultrasonic.
|
a. |
Calibration
shall include the complete ultrasonic examination system. The original
calibration must be performed on the basic calibration block. Check shall be
made to verify the distance range calibration and distance amplitude
correction, check shall include the entire examination system.
|
|||||||||
b.
|
Pada
semua kalibrasi, sangat penting diperhatikan bahwa maximum indikasi
dihasilkan oleh gelombang yang tegak lurus dengan lubang samping dan kowakan
(notch). Pusat dari probe harus minimum 38 mm dari sisi terdekat blok,
(menggerakkan gelombang ke arah pojok perpotongan lubang samping dengan
sisi blok mungkin akan menghasilkan amplitudo yang tinggi dan beam path yang
panjang; beam path ini tidak boleh dipakai untuk kalibrasi). Untuk pengujian
sistem kontak, beda temperatur antara permukaan benda yang di uji dan blok kalibrasi harus
tidak lebih dari 14°C.
|
b. |
In
all calibration, it is important that maximum indications be obtained with
the sound beam oriented perpendicular to the exits of the side drilled hole
and notches. The center line of the search unit shall be at least 38 mm from
the nearest side of the block. (rotation of the beam into the corner
formed by the hole and the side of the block may produce a higher amplitude
at longer-beam path; this beam path shall not be used for calibration ). For
contact examination, different the temperature of the examination and basic
calibration block surface shall be within 14°C.
|
|||||||||
|
|
|
|
|||||||||
9.2.2
|
Teknik
Kalibrasi probe sudut
|
9.2.2 |
Angel Beam
Calibration technique
|
|||||||||
|
Minimum kalibrasi harus meliputi
pengukuran 2 sebagai berikut.
|
|
The minimum calibration shall
provide the following measurements.
|
|||||||||
9.2.2.1
|
Kalibrasi jarak (lihat Fig. 2)
|
9.2.2.1 |
Distance range calibration (see Fig.
2)
|
|||||||||
|
Kalibrasi
range jarak harus dilaksanakan sesuai dengan ASME V, article 4 (B – 410) atau
dengan menggunakan permukaan radius dengan IIW V1 atau V2. Atur display
kalibrasi nol dan range jarak untuk metal yang diinginkan untuk pengujian
(Fig. 2).
|
|
Distance
range calibration shall be performed as per ASME Section V, article 4 (B –
410) or by using reflection from concentric cylindrical surface provide by
IIW V1 or V2. Adjust display zero and distance range for metal path
calibration so as to be appropriate for examination (Fig. 2).
|
|||||||||
9.2.2.2
|
Koreksi Jarak Amplitudo
|
9.2.2.2 |
Distance-Amplitude Correction
|
|||||||||
a.
|
Posisikan probe (lihat Fig. 3) untuk maximum respond:
-
dari 1½ t SDH
untuk tebal material t < 25
mm
-
dari ¼ atau ¾
t SDH untuk material t >
25 mm
SDH diatas dinamakan reference SDH.
Referensi lain bisa dipilih selain yang disebut diatas,
untuk mencakup semua pengujian.
|
a. |
Position the search unit (see Fig 3) for maximum
response:
-
from the 1½ t SDH
for material thickness t < 25 mm
-
from ¼ or ¾ t SDH for material thickness t > 25 mm
The above SDH is called as reference SDH.
The reference SDH may be selected other than mentioned
above, to provide the coverage testing
|
|||||||||
b.
|
Atur sensitivity control
untuk menghasilkan indikasi 80% (±5% tinggi full screen)
dari reference SDH.
Tandai ujung indikasi
pada screen dengan pensil tahan grease atau dengan marker yang cocok.
Catat dB dari reference
SDH, kemudian dB ini adalah sebagai reference level.
|
b. |
Adjust the sensitivity
control to provide 80% (±5% of full screen height) indication from the
reference SDH.
Mark the peak of the
indication on the screen with a grease pencil or other suitable marker.
Record the dB from the
reference SDH, the dB is considered as reference level.
|
|||||||||
c.
|
Atur
sensitivity control untuk menghasilkan 40% of full screen height atau 50 %
DAC (±5%),
dengan mengurangi 6 db dari reference level.
Tandai
ujung 2 indikasi pada screen.
|
c. |
Adjusting the
sensitivity control to provide 40% of full screen height or 50 % DAC (±5%),
by reducing 6 dB from the reference level.
Mark the peak of
the indications on the screen.
|
|||||||||
d.
|
Atur
sensitivity control untuk menghasilkan 16% tinggi full screen atau 20 % DAC (±5%),
dengan mengurangi 14 db dari reference level.
Tandai
ujung 2 indikasi pada screen.
|
d. |
Adjusting the
sensitivity control to provide 16% of full screen height or 20 % DAC (±5%),
by reducing 14 dB from the reference level.
Mark the peak of
the indications on the screen.
|
|||||||||
e.
|
Tanpa
merubah Tombol range, atur sensitivity control ke posisi reference level,
kemudian arahkan probe ke SDH 2 lainnya untuk menghasilkan maximum
respond.
a.
|
e. |
Without changing
the Range Control adjustment, adjusting the sensitivity control to the reference
level, then position the search unit for maximum response from another
SDHs.
a.
|
|||||||||
f.
|
Ulangi step (c)
dan (d) diatas. Hubungkan tanda – tanda yang ada di screen respond dari SDH 2
tersebut untuk menghasilkan distance amplitude curve
|
f. |
Repeat
the step (c) and (d) above. Connect the screen marks respond from the side
drilled SDHs to provide the distance amplitude curve (DAC).
|
|||||||||
9.2.2.3
|
Pengukuran amplitudo echo dari notch
permukaan pada blok kalibrasi (lihat Fig. 4)
|
9.2.2.3 |
Echo amplitude measurement from the
surface notch in the basic calibration block (see Fig. 4)
|
|||||||||
a.
|
Posisikan probe untuk menghasilkan
amplitudo maximum dari notch kotak di permukaan dibaliknya, tandai “X“ pada
ujung indikasi pada screen dari notch tersebut.
|
a. |
1.
Position the search unit for maximum
amplitude from the square notch on the opposite surface, “X” mark the peak of
the indication on the screen from the notch.
|
|||||||||
b.
|
Notch kotak di
permukaan dibaliknya mungkin bisa memberi indikasi 2 banding 1 diatas DAC
dari probe 450 and 1/2 DAC at probe 60 0. Oleh karena
itu, indikasi dari notch kotak harus diperhatikan jika evaluasi reflektor 2
pada posisi permukaan dibaliknya.
|
b. |
2. The
opposite surface square notch may give an indication 2 to 1 above DAC at 450
probe and 1/2 DAC at 60 0 probe.
Therefore, the indication from the square notch must be consider when
evaluating reflectors at the opposite surface.
|
|||||||||
|
|
|
|
|||||||||
9.2.3
|
Kalibrasi probe normal.
|
9.2.3 |
Straight beam calibration.
|
|||||||||
|
Kalibrasi harus meliputi pengukuran2 sebagai
berikut
|
|
The calibration shall provide the following measurements.
|
|||||||||
9.2.3.1
|
Kalibrasi jarak (lihat Fig. 5, sebagai contoh)
Kalibrasi dari pengukuran beam path bisa digunakan
control range dengan memposisikan refleksi dari backwall blok untuk pembagian divisi yang sama untuk tebal yang diukur dari blok
IIW V1 atau V2 atau blok standar yang setara.
|
9.2.3.1 |
Distance range calibration (see Fig.
5, for example)
Calibration
of beam path measurement may use range control by positioning of the block
backwall reflection to the distance division number (or multiple) equal to
the measured thickness of IIW VI or
V2 or equivalent standard test block.
|
|||||||||
|
|
|
|
|||||||||
|
|
|
|
|||||||||
|
|
|
|
|||||||||
|
|
|
|
|||||||||
9.2.4
|
Konfirmasi Kalibrasi
|
9.2.4 |
Calibration Confirmation
|
|||||||||
a. |
Tahapan Konfirmasi kalibrasi
|
a. |
System Changes.
|
|||||||||
|
Check
kalibarasi harus dilaksanakan, jika ada penggantian part dari system, untuk
memenuhi kalibrasi jarak dan seting sensitivity seperti par. 9.2.4. c (1),
(2), sbagai berikut :
|
|
Calibration
checks shall be performed, when any part of the examination system is
changed, to verify the distance range points and sensitivity settings satisfy
of par. 9.2.4. c (1), (2), the followings:
|
|||||||||
1. |
Kalibrasi
check harus dilaksanakan pada blok kalibrasi (V1 atau V2) untuk verifikasi kalibrasi
jarak
|
1. |
calibration
check shall be made on the calibration block (V1 or V2) verify that distance
range points
|
|||||||||
2. |
Kalibrasi
check harus dilaksanakan pada basic Calibration block (BCB) untuk verifikasi
setting sensitivity
|
2. |
calibration
check shall be made on the basic calibration block (B C B) to verify the sensitivity setting
|
|||||||||
b. |
Periodic
Examination Check
|
b. |
Periodic
Examination Check
|
|||||||||
|
Kalibrasi
check harus dilaksanakan pada minimum satu reflector didalam Basic Calibration Bock pada kondisi berikut
:
|
|
A
calibration check shall be made on at least one of the basic reflectors in
the Basic Calibration Block of the
following conditions :
|
|||||||||
1. |
a. Pada setiap akhir dari dari pengujian atau
|
1. |
a. At the finish of each examination or
|
|||||||||
2. |
b. Pengujian yang serupa secara seri.
|
2. |
b. Series of similar examination.
|
|||||||||
3. |
c. Jika personel yang melakukan pengujian diganti.
|
3. |
c. When examination personell are changed.
|
|||||||||
c. |
d. Confirmation Acceptance Values
|
c. |
Confirmation Acceptance Values
|
|||||||||
|
e. Semua indikasi yang direcord sejak kalibrasi terakhir
yang valid atau check kalibrasi harus di uji ulang dengan kalibrasi
yang benar dan nilai pada data sheetnya harus dirubah, jika:
|
|
d. All recorded indication since the last valid
calibration or calibration check shall be re examined with the
corrected calibration and the values shall be changed on their data sheets,
if :
|
|||||||||
1.
|
f.
setiap poin pada
curva DAC telah berubah dari garis
kalibrasi jarak lebih besar dari 10%
dari pembacaan jarak atau 5% dari jarak penuh (screen), maka diambil yang
lebih besar, koreksi jarak kalibrasi dan beri catatan dari koreksi pada
record pengujian
|
1. |
e. a point on the DAC curve has moved
on distance line more than 10 % of distance reading or 5% of full
distance, whichever is greater, correct the distance range calibration and note
the correction in the examination record or
|
|||||||||
2.
|
g. setting sensitivity telah mengalami penurunan atau
kenaikan lebih dari 20% atau 2 db dari
amplitudo nya
|
2. |
f.
any sensitivity
setting has decreased or increased by more than 20% or 2 dB of its amplitude
|
|||||||||
|
|
|
|
|||||||||
9.2.5
|
Pengujian
|
9.2.5 |
Examination
|
|||||||||
9.2.5.1
|
Ruang lingkup
|
9.2.5.1 |
Coverage
|
|||||||||
|
Volume harus di uji dengan menggerakkan probe keseluruh
permukaan yang di uji sehingga bisa menscan seluruh volume pengujian. Setiap
jalan / jalur dari probe harus overlaping minimum 10 % dari ukuran probe,
tegak lurus dengan dimensi dari arah scaning.
|
|
The volume shall be examined by moving the search unit
over the examination surface so as to scan the entire examination volume.
Each pass the search unit shall be overlap a minimum of 10% of transducer
(element), dimension perpendicular to the direction of the scan.
|
|||||||||
9.2.5.2
|
Syarat 2
Scanning
|
9.2.5.2 |
Scanning
Requirements
|
|||||||||
a.
|
Kecepatan dari manual scaning harus tidak lebih dari 6
in / sec. (150 mm/sec.)
|
a. |
The rate of manual scanning shall not exceed 6 in/sec
(150 mm/sec).
|
|||||||||
b.
|
Scaning
harus dilaksanakan dengan mengatur gain seting minimum dua kali dari primary
reference level. Evaluasi harus dilaksanakan dengan mengacu pada primary
reference level.
|
b. |
The
scanning shall be performed at gain setting at least two times the primary
reference level. Evaluation shall be performed with respect to the primary
reference level.
|
|||||||||
c.
|
Scaning harus dilaksanakan dari
kedua sisi lasan (jika memungkinkan).
|
c. |
The scanning shall be performed from
both sides of the weld (if the accessible)
|
|||||||||
|
|
|
|
|||||||||
9.2.5.3
|
Scaning dengan
Probe Normal
|
9.2.5.3 |
Straight Beam
Scanning
|
|||||||||
a.
|
Scaning
probe normal harus dilaksanakan pada volume dari base material dimana akan
dilewati probe sudut untuk mengetahui reflector yang mungkin membatasi
kemampuan probe sudut untuk menguji volume lasan, dan ini tidak digunakan
sebagai acceptance criteria.
|
a. |
The
straight beam scanning shall be performed on the volume of base material
through which the angle beams will travel to locate any reflectors that might
limit the ability of the angle beam to examine the weld volume , and is not
to be used for an acceptance - rejection examination.
|
|||||||||
b.
|
Lokasi dan area yang ditemukan reflektor harus direcord
dan menjadi bahan pertimbangan Engineering.
|
b. |
Location and areas of such reflectors found shall be
recorded and subject to Engineering evaluation.
|
|||||||||
c.
|
Scaning probe normal harus dilaksanakan pada volume
dari base material dimana cacat mungkin mengakibatkan gangguan di service
yang di refer oleh Code section lainnya atau persyaratan lainnya.
|
c. |
The straight beam scanning shall be performed on the
volume of base material in which the defects may affect the service
referenced by other Code section or other requirements.
|
|||||||||
9.2.5.4
|
Scanning dengan
probe sudut untuk reflektor yang paralel dengan lasan.
|
9.2.5.4 |
Angle beam
scanning for reflectors parallel to the weld
|
|||||||||
|
Probe sudut harus diarahkan ke lasan dengan sudut yang
benar kearah axis dari lasan dari daerah yang memungkinkan. Probe harus di
manupulasi sehingga semua energi ultrasonic berjalan melalui volume yang
diperlukan dari lasan dan base metal sekitarnya.
|
|
The angle beam shall be directed at approximate right
angle to the weld axis from to directions where possible. The search unit
shall be manipulated so that the ultrasonic energy pass trough the required
volumes of weld and adjacent base metal.
|
|||||||||
9.2.5.5
|
Scanning dengan probe sudut untuk
reflektor yang melintang dengan lasan.
|
9.2.5.5 |
Angle beam scanning for reflection
oriented transversal to the weld.
|
|||||||||
a.
|
Probe
sudut harus diarahkan sedapat mungkin paralel dengan axis dari lasan. Probe
harus di manupulasi sehingga semua energy ultrasonic berjalan melalui volume
yang diperlukan dari lasan dan base metal sekitarnya.
|
a. |
The
angle beam shall be directed essentially parallel to the weld axis. The
search unit shall be manipulated so that the angle beam through the required
volumes of weld and adjacent base metal.
|
|||||||||
b.
|
Probe harus di putar 180 derajat dan pengujian
diulangi.
|
b. |
The search unit shall be rotated 180 deg. and the
examination repeated.
|
|||||||||
c.
|
Probe sudut bisa digunakan untuk menguji base
materials, probe harus diarahkan ke
segala arah untuk scanning cacat yang arahnya bermacam – macam.
|
c. |
The angle beam may be used for examination base
materials, the search unit shall be manipulated in any direction for scanning
the any orientation defects.
|
|||||||||
|
|
|
|
|||||||||
9.2.6
|
Evaluasi
ukuran cacat.
|
9.2.6 |
Discontinuity
size evaluation
|
|||||||||
9.2.6.1
|
Tes dengan
Probe normal.
|
9.2.6.1 |
Straight
Beam Testing
|
|||||||||
a.
|
Jika cacat yang terjadi lebih kecil dari ukuran
probe, maka evaluasi dari ukuran cacat harus dilaksanakan dengan memulai dari
arah luar dan kedalam dari cacat, menggerakkan probe kearah area cacat sampai
terdapat indikasi pada layar kelihatan. Ujung dari probe/search unit diindikasikan
sebagai ujung dari cacatnya.
|
a. |
When discontinuity is smaller than the transducer
size, the size evaluation shall be made by beginning outside of the
discontinuity, moving the transducer toward the area of discontinuity until
an indication on the display begins to form. The leading edge of the search
unit at this point is indicative of the edge of the discontinuity
|
|||||||||
b.
|
Jika cacat lebih besar dari ukuran probe, kehilangan
refleksi back wall akan terjadi dan evaluasi metode dengan menggunakan cara 6
db. drop (50% penurunan dari ketinggian amplitudo aslinya) harus
digunakan dan untuk menentukan ujung
cacat dengan mengukur center line dari probe.
|
b. |
When discontinuity is larger than the transducer
size, a full loss of back reflection will occur and evaluation method of a 6
dB loss of amplitude (drop to
50 % from the original screen height) shall be used and measurement to the
center line of the transducer is determining discontinuity edge.
|
|||||||||
9.2.6.2
|
Tes dengan
Probe sudut.
|
9.2.6.2 |
Angle Beam
Testing
|
|||||||||
a.
|
Panjang dan tinggi dari cacat harus ditentukan dengan
menggerakkan probe dari daerah cacat sampai ada penurunan amplitudo 6 db.
drop (turun 50% dari amplitudo aslinya) dan untuk menentukan ujung cacat
dengan mengukur center line dari probe.
|
a. |
The length and height of
the discontinuity shall be determined by moving the transducer from the
discontinuity area until a 6 dB loss of amplitude (drop to 50 % from the
original screen height) and measurement to the center line of the transducer
is determining discontinuity edge
|
|||||||||
b.
|
Reflektor yang berbentuk bulat bisa di bedakan dengan
yang berbentuk planar dengan melihat respond yang ditunjukkan pada layar
seperti akan begitu mencolok drop dari amplitudonya jika probe dijauhkan dari
area cacat atau dimanipulasi dengan sudut tertentu dari maximum respond dari
area cacat.
|
b. |
The spherical reflector may be differentiated from
Planar reflector by their respond
shown as amplitude significant drop when the transducer moving away from the
indicative area or when the transducer manipulated to an angle from the maximum respond of
indicative area.
|
|||||||||
|
|
|
|
|||||||||
9.2.7
|
Acceptance
standard
|
9.2.7 |
Acceptance
standards
|
|||||||||
9.2.7.1
|
Semua
indikasi yang mengahsilkan amplitudo lebih besar dari 20% dari the reference
level harus di investigasi oleh operator untuk menentukan bentuk,
identifikasi dan lokasi dari indikasi tersebut dan mengevaluasi dengan
mengacu pada acceptance standard berikut ini:
|
9.2.7.1 |
All
imperfections that produce an amplitude greater than 20% of the reference
level shall be investigated to the extend that the operator can determine the
shape, identity and location of all such imperfections and evaluate them in
terms of the following acceptance
standards :
|
|||||||||
9.2.7.2
|
Indikasi yang termasuk kategori crack, lack of fusion,
atau incomplete penetration semuanya tidak diterima.
|
9.2.7.2 |
Indication characterized as cracks, lack of fusion,
or incomplete penetration are unacceptable regardless of length.
|
|||||||||
9.2.7.3
|
Indikasi
lain tidak diterima jika melebihi amplitudo reference level dan mempunyai panjang melebihi :
|
9.2.7.3 |
Other
imperfection are unacceptable if the indications exceed the reference level amplitude and have length unit exceed
:
|
|||||||||
|
6.0
mm untuk T up to 19.0 mm.
1/3
T for T untuk 19.0 to 57.0 mm.
19.0 mm untuk T over 57.0
mm.
|
|
6.0
mm for T up to 19.0 mm.
1/3
T for T from 19.0 to 57.0 mm.
19.0 mm for T over 57.0
mm.
|
|||||||||
9.2.7.4
|
T adalah tebal dari lasan tanpa ditambah
reinforcement. Untuk suatu butt join
yang terdiri dari dua ketebalan yang berbeda, T adalah tebal dari bagian
yang tertipis. Untuk lasan fillet tambahan dari groove weld, ketebalan
throat dari fillet harus dimasukkan dalam penghitungan T.
|
9.2.7.4 |
T is the thickness at the weld excluding any
allowable reinforcement. For butt weld joining two members having different
thickness at the weld, T is the thinner of these two thicknesses. For fillet
weld in addition of the groove welds, the thickness of the throat of
the fillet shall be included in T.
|
|||||||||
9.2.7.4
|
Untuk API 650, acceptante standard harus didasarkan
persetujuan antara pembeli dan pabrikan.
|
9.2.7.5 |
For API 650,
Acceptance standards shall be agreed upon by the purchaser and the
manufacturer.
|
|||||||||
|
|
|
|
|||||||||
9.2.8
|
Pengujian hasil
perbaikan
|
9.2.8 |
Examination of
repairs
|
|||||||||
|
Cacat
yang tidak diterima harus di buang dan di repair. Repair harus di uji ulang
dengan prosedure yang sama yang digunakan untuk cacat aslinya.
|
|
Unacceptable
discontinuities shall be removed and repaired. Repairs shall be reexamined by
the same procedures used for original of the discontinuity.
|