Rabu, 12 April 2017

contoh prosedur aws untuk ultrasonic testing


         Daftar ISI  /  TABLE OF CONTENT

 Page no.
1
Tujuan
1
Purpose
4
2
Ruang Lingkup
2
Scope
4
3
Laporan dan Catatan
3
Records
4
4
Documen Referensi
4
Reference Documents
5
5
Definisi – Definisi
5
Definitions
5
6
Kualifikasi dan Tanggung Jawab
6
Qualification and Responsibility
5
7
Perlengkapan
7
Tools, Gages, Fixtures
6
7.1
Alat – alat  Ultrasonic
7.1
Ultrasonic Equipments
6
7.1.2
Linearitas Horizontal
7.1.2
Horizontal Linearity
6
7.1.3
Syarat 2 untuk Pengujian Instrumen
7.1.3
Requirement for Test Instruments.
6
7.1.4
Kalibrasi dari Pengujian Instrumen
7.1.4
Calibration of Test Instruments.
6
7.2
Search Units.
7.2
Search Units.
7
7.2.1
Search Units untuk Probe Normal (Longitudinal Wave)
7.2.1
Straight-Beam (Longitudinal Wave) Search Units.
7
7.2.2
Resolusi untuk Probe Normal.
7.2.2
Resolution for Straight-Beam.
7
7.2.3
Search Units Probe Sudut.
7.2.3
Angle-Beam Search Units.
7
7.2.4
Frekuensi.
7.2.4
Frequency.
7
7.2.5
Dimensi Transducer .
7.2.5
Transducer Dimensions.
7
7.2.6
Sudut.
7.2.6
Angles.
7
7.2.7
Resolusi untuk Probe Sudut.
7.2.7
Resolution for Angle-Beam.
8
7.2.8
Penandaan
7.2.8
Marking.
8
7.2.9
Jarak Ujung Probe
7.2.9
Edge Distance.
8
7.3
Standar Acuan
7.3
Reference Standards.
8
8
Instruksi – instruksi
8
Instructions
8
8.1
Prosedur Kalibrasi dan Kualifikasi   Peralatan
8.1
Equipment Calibration and Qualification Procedure
8
8.1.1
Linearitas Horizontal.
8.1.1
Horizontal Linearity.
8
8.1.2
Gain Control (dB Accuracy ).
8.1.2
Gain Control (dB Accuracy ).
9
8.1.3
Rumus Decibel.
8.1.3
Decibel Equation.
12
8.1.4
Prosedure Refleksi Internal.
8.1.4
Internal Reflections Procedure.
12
8.2
Kalibrasi dari Search Units.
8.2
Calibration of Search Units.
13
8.2.1
Kalibrasi dari Search Units probe sudut.
8.2.1
Calibration of Angle-Beam Search Units.
13
8.2.2
Titik Index.
8.2.2
Index point.
13
8.2.3
Sudut
8.2.3
Angles.
13
8.2.4
Persyaratan untuk resolusi dari probe sudut .
8.2.4
Angle Beam Resolution Requirements.
14
8.2.5
Kalibrasi dari Search Units Probe Normal (Longitudinal Mode ).
8.2.5
Calibration of Straight-Beam Search Units (Longitudinal Mode).
15
8.2.6
Persyaratan Resolusi Probe Normal
8.2.6
Straight-Beam Resolution Requirement
15
8.2.7
Kualifikasi Linearitas Horizontal.
8.2.7
Horizontal Linearity Qualification
15
8.2.8
Kualifikasi Gain Control (Attenuation).
8.2.8
Gain Control (Attenuation) Qualification.
16
8.3
Kalibrasi untuk  Pengujian
8.3
Calibration for Testing
16
8.3.1
Posisi Kontrol Reject.
8.3.1
Position of Reject Control.

16
8.3.2
Reflector yang dilarang.
8.3.2
Prohibited Reflectors.
16
8.3.3
Teknik.
8.3.3
Technique.
16
8.3.4
Kalibrasi Ulang.
8.3.4
Recalibration.
16
8.4
Pengujian Probe Normal untuk  Base Metal.
8.4
Straight-Beam Testing of Base Metal.
16
8.5
Kalibrasi untuk Pengujian Probe Sudut.
8.5
Calibration for Angle-Beam Testing.
17
8.5.1
Horizontal Sweep.
8.5.1
Horizontal Sweep.
17
8.5.2
Zero Reference Level.
8.5.2
Zero Reference Level.
17
8.6
Prosedur Pengujian.
8.6
Testing Procedures
18
8.6.1
“X” Line.
8.6.1
“X” Line.
18
8.6.2
“Y” Line.
8.6.2
“Y” Line.
18
8.6.3
Kebersihan.
8.6.3
Cleanliness.
18
8.6.4
Couplants.
8.6.4
Couplants.
18
8.6.5
Maksud dari pengujian.
8.6.5
Extent of Testing.
19
8.6.6
Kesulitan Akses.
8.6.6
Inaccessibility.
19
8.7
Pengujian Lasan.
8.7
Testing of Welds.
20
8.8
Scanning.
8.8
Scanning.
20
8.8.1
Longitudinal Discontinuities
8.8.1
Longitudinal Discontinuities
20
8.8.2
Transverse Discontinuities
8.8.2
Transverse Discontinuities
20
8.9
Butt Joints.
8.9
Butt Joints.
21
9
Prosedure Evaluasi  Ukuran Discontinuities
9
Discontinuity Size Evaluation Procedures
22
9.1
Pengujian dengan Probe Normal (Longitudinal)
9.1
Straight-Beam (Longitudinal) Testing.
22
9.2
Pengujian dengan Probe Sudut (Shear).
9.2
Angle-Beam (Shear) Testing.
22
10
Kriteria Penerimaan
10
Acceptance Criteria
23
10.4
Kriteria Penerimaan untuk Statically Loaded  Nontubular Connections.
10.4
Acceptance Criteria for Statically Loaded  Nontubular Connections.
24
10.5
Kriteria Penerimaan untuk Cyclically Loaded Nontubular Connections.
10.5
Acceptance Criteria for Cyclically Loaded Nontubular Connections.
24
11
Identifikasi pada Area yang di Reject.
11
Identification of Rejected Area.
25
12
Perbaikan.
12
Repair.
25
13
Laporan Pengujian Ulang.
13
Retest Reports.
25
14
Steel Backing.
14
Steel Backing.
26
15
Lampiran
15
Attachments
27~ 40






1
Tujuan
1
Purpose
a.
Prosedur ini dipakai untuk petunjuk sewaktu melakukan Pengujian Ultrasonic Kontak Langsung untuk Structural Steel welds dan HAZ(s), menggunakan pulse echo type.
a.
This  procedure provides guidance during the conduct Direct Contact Ultrasonic Examination of Structural Steel welds and HAZ(s), using pulse echo type.
b.

Prosedur ini terdiri dari versi Inggris dan Indonesia, jika ada konflik antara keduanya maka versi Inggris yang dipakai

b.
This procedure consists of English and Indonesian versions, should there is conflict between these versions the English version shall govern.




2
Ruang Lingkup
2
Scope
2.1
Prosedur ini berisi syarat –syarat dan acceptance standards Pengujian Ultrasonic Kontak Langsung untuk Structural Steel full penetration groove welds dari ferrous material (carbon steel, low alloy steel and stainless steel material) dengan  thickness antara  0.3125 inch ( 8.0 mm ) sampai 8.0 inch ( 200 mm ).
2.1
This procedure provides the requirements and acceptance standards for the contact ultrasonic examination for full penetration groove welds of ferrous material (carbon steel, low alloy steel and stainless steel material) with thickness between 0.3125 inch ( 8.0 mm ) upto 8.0 inch ( 200 mm ).
2.2
Prosedur ini tidak dimaksudkan untuk pelaksanaan pengujian ultrasonic pada sambungan terbuat dari pipa.
2.2
This procedure is not intended for ultrasonic examination of  tubular connections.




3
Laporan dan Catatan
3
Records
3.1
Hasil Pengujian Ultrasonic harus ditulis di Ultrasonic Test Report AWS Form               (attached) atau Form yang typical.
3.1
Ultrasonic Examination Results shall be reported in the provided “ Ultrasonic Test Report AWS Form “ (attached) or typical form.
3.2
Isi dari laporan harus mempunyai cukup informasi, jelas identifikasi partnya, nomor jointnya dan area yang diuji.
3.2
Content of reports shall have suffient informations, clearly identify parts, joint numbers and area of examinations.
3.3
Sebagai tambahan diatas, report untuk lasan yang acceptable perlu cukup berisi informasi yang mengenai nomor joint, tanda tangan operator, dan keberterimaan dari lasan.
3.3
In addition of the above, the report form  for welds that are acceptable need only contain sufficient information to identify the weld joint numbers , the operator signature, and acceptability of the welds.
3.4
Sebagai tambahan diatas, report untuk lasan yang tidak diterima harus juga menunjukkan quality sebelum direpair dan bisa direview oleh orang – orang yang terkait.
3.4
In addition of the above, the report form for welds that are unacceptable shall show unacceptable quality prior to repair and shall be subject to be  reviewed by  involving party.
3.5
Jika Customer tidak minta Report UT untuk dikirim, Report UT harus disimpan sedikitnya satu tahun penuh setelah pekerjaan komplet selesai .
3.5
Unless requested by Customer for delivering the completed UT reports, the UT reports shall be retained at least one full year after completion of
works.




4
Dokumen Referensi
4
Reference Documents
4.1
 Written Practice Procedure No.:SMS-601-P-001.
4.1
 Written Practice Procedure No.:SMS-601-P-001.
4.2
A W S D 1.1 – 2006 Edition ;  Section 6 Part C ( 6.13 ), Part F .
4.2
A W S D 1.1 – 2006 Edition ;  Section 6 Part C ( 6.13 ), Part F .




5
Definisi – Definisi
5
Definitions
5.1
Zero reference level : respond dari ultrasonic dari refector 0.060” (1.59 mm)   di blok kalibrasi IIW - V1, secara electronic di atur sampai persen tertentu  dari ketinggian screen.
5.1
Zero reference level  : the ultrasonic response from the 0.060” (1.59 mm)  reflector in the IIW – V1 Standard Block Calibration, electronically adjusted to a specified percentage  of  the full screen height.
5.2
Search Unit : sebuah alat electro acoustic digunakan untuk meneruskan atau menerima ultrasonic energy.
5.2
Search Unit : an Electro acoustic  device used to transmit or receive ultrasonic energy.
5.3
Kalibrasi : hubungan dari respon sistem ultrasonic  dengan reflektor kalibrasi.
5.3
Calibration : correlation of the ultrasonic system response with calibration reflector (s).
5.4
Transducer ( Probe ) : sebuah alat electro acoustical untuk merubah energi listrik ke  energi getaran dan sebaliknya.
5.4
Transducer ( Probe ) : an electroacoustical device for converting electrical energy into acoustical energy and vice versa.
5.5
Probe Normal : sebuah gelombang getaran pulsa yang berjalan normal (tegak lurus) dengan permukaan yang di tes.
5.5
Straight beam : a vibrating pulse wave train travelling normal to the test surface.
5.6
Probe Sudut : sebuah terminologi yang digunakan untuk menyatakan sebuah sudut datang atau pembengkokan selain normal (tegak lurus) dengan permukaan yang di tes, seperti sudut pengujian, sudut probe, sudut beam gelombang longitudinal, dan gelombang beam shear.
5.6
Angle beam : a term used to describe an angle of incidence or refraction other than normal to the surface of the test object, as an angle beam examination, angle beam search unit, angle  beam longitudinal wave, and angle beam shear waves.




6
Kualifikasi dan Tanggung Jawab
6
Qualification and Responsibility
6.1
Personel yang melaksanakan pengujian sesuai dengan requiremen prosedure ini harus diqualifikasi dan disertifikasi sesuai dengan PT. SPEKTRA MEGAH SEMESTA’s Written Practice, no. SMS-601-P-001..
6.1
Personnel performing examinations to the requirements this procedure shall be qualified and certified in accordance PT. SPEKTRA MEGAH SEMESTA’s Written Practice, no. SMS-601-P-001.
6.2
Ujian untuk qualifikasi  harus termasuk demonstrasi kemampuan UT operator dalam pengaplikasian aturan di Code yang sudah di adopt dan dimasukkan kedalam prosedure ini, dalam mendeteksi dan mendesposisi cacat – cacat.
6.2
The examination for qualification shall include UT operator demonstration his ability to apply the rules of the Code which has been adopted and included in  this procedure, in accurate detection and disposition of discontinuitis.
6.3
Personel yang melaksanakan tes dan mengevaluasi dengan prosedure ini harus memiliki Level II atau Level III.
6.3
Personnel performing examinations and evaluation to the requirements of this procedure shall be Level II or Level III




7
Perlengkapan
7
Tools, Gages, Fixtures
7.1
Alat – alat  Ultrasonic
7.1
Ultrasonic Equipments
7.1.1
UT instrument harus pulse echo type cocok untuk digunakan dengan transducer dengan oscilasi pada frekwensi antara 1 MHz to 6 MHz. Display harus “A“ scan rectified video trace. Dinamic range display instrument harus sedemikian rupa sehingga dengan mudah perbedaan dari 1 db amplitude dengan mudah bisa dideteksi di display.
UT instrument yang operasi frekwensinya diluar range tersebut mungkin bisa digunakan jika sensitivitinya  sama atau lebih bagus bisa demonstrasikan dan didokumentasikan.
7.1.1
The UT instrument shall be the pulse-echo type suitable for use with transducer oscillating at frequencies between 1 MHz to 6 MHz. Display shall be “A“ scan rectified video trace. The dynamic range of the instrument’s display shall be such that a difference of 1 dB of amplitude can be easily detected on the display.

The UT instrument operating at other frequencies may be used if equal or better sensitivity can be demonstrated and documented
7.1.2
Linearitas Horizontal
7.1.2
Horizontal Linearity

Lineariti horizontal dari instrument harus diqualifikasi seluruh sound path distance yang digunakan, prosedur ini akan didetail pada paragraf 8.1.1.

The horizontal linearity of the instrument shall be qualified over the full sound path distance to be used in testing , the procedure are detailed in par. 8.1.1.
7.1.3
Syarat 2 untuk pengujian Instrumen
7.1.3
Requirement for Test Instruments.

Test Instrument harus termasuk internal stabilisasi sehingga setelah pemanasan, tidak akan terjadi ada respon variasi lebih besar dari ± 1 dB dengan perubahan suply voltage 15% dari nominal atau, terjadi masalah dengan  baterey, selama perubahan lama operasinya. Harus ada sebuah alarm atau signal jika voltage baterey drop sebelum instrument mati dikarenakan baterey habis.

Test instruments shall include internal stabilization so that after warm-up, no variation in response greater than ± 1 dB occurs with a supply voltage change of 15% nominal or, in the case of a battery, throughout the charge operating life. There shall be an alarm or meter to signal a drop in battery voltage prior to instrument shutoff due to battery exhaustion.

7.1.4
Kalibrasi dari Pengujian Instrumen.
7.1.4
Calibration of Test Instruments.

Test Instrument harus mempunyai gain kontrol (attenuator) yang terkalibrasi yang bisa di atur pada setiap step 1 atau 2 dB seluruh range dan paling sedikit 60 dB. Keakuratan dari seting attenuator harus dalam range plus atau minus 1 dB. Prosedur ini akan didetail pada paragraf 8.1.2.

The test instrument shall have a calibrated gain control (attenuator) adjustable in discrete 1 or 2 dB steps over a range of at least 60 dB. The accuracy of the attenuator settings shall be within plus or minus 1 dB. The procedure are detailed in par. 8.1.2.




7.2
Search Units.
7.2
Search Units.
7.2.1
Search Units untuk Probe Normal (Longitudinal Wave)
7.2.1
Straight-Beam (Longitudinal Wave) Search Units.

Straight-beam (longitudinal wave) search unit transducers harus mempunyai aktive area tidak kurang dari ½  inch persegi (323 mm persegi) atau tidak lebih dari 1 inch persegi (645 mm persegi). Tranducer harus bundar atau persegi.

Straight-beam (longitudinal wave) search unit transducers shall have an active area of not less than 1/2 square inches [323 square millimeters] nor more than 1 square inch [645 square millimeters]. The transducer shall be round or square
7.2.2
Resolusi untuk Probe Normal.
7.2.2
Resolution for Straight-Beam.

Transducers harus mampu membedakan tiga reflektor, prosedur ini akan didetail pada paragraf 8.2.6.

Transducers shall be capable of resolving the three reflections, the procedure are detailed in par. 8.2.6.




7.2.3
Search Units Probe Sudut.
7.2.3
Angle-Beam Search Units.

Probe Sudut  harus terdiri dari sebuah tranducer dan wedge yang bersudut.
Unit bisa terpisah maupun menjadi satu kesatuan jadi satu unit.

Angle-beam search units shall consist of a transducer and an angle wedge.
The unit may be comprised of the two separate elements or may be an integral unit.
7.2.4
Frequency.
7.2.4
Frequency.

Frequensi probe harus antara 2 and 2.5 MHz.

The transducer frequency shall be between 2 and 2.5 MHz, inclusive.
7.2.5
Dimensi Transducer.
7.2.5
Transducer Dimensions.

Crystal probe harus berbentuk persegi atau persegi panjang dan bervariasi dari 5/8 in. sampai 1 in. [15 sampai 25 mm] lebar dan dari 5/8 sampai 13/16 in. [15 sampai 20 mm] tinggi (lihat Figure 6.20).
Perbandingan antara lebar dan panjang maximum harus diantara 1.2 ke 1.0, dan lebar dan panjang minimum harus diantara 1.0 ke 1.0.

The transducer crystal shall be square or rectangular in shape and may vary from 5/8 in. to 1 in. [15 to 25 mm] in width and from 5/8 to 13/16 in. [15 to 20 mm] in height (see Figure 6.20).
The maximum width to height ratio shall be 1.2 to 1.0, and the minimum width-to-height ratio shall be 1.0 to 1.0.
7.2.6
Sudut.
7.2.6
Angles.

Search unit harus menghasilkan sound beam dalam material yang ditest dalam range plus atau minus 20 dari salah satu sudut berikut : 70°, 60°, atau 45.

The search unit shall produce a sound beam in the material being tested within plus or minus 2° of one of the following proper angles: 70°, 60°, or 45.
7.2.7
Resolusi untuk Probe sudut.
7.2.7
Resolution for Angle-Beam.

Transducers harus mampu membedakan tiga reflektor, prosedur ini akan didetail pada paragraf 8.2.4.

Transducers shall be capable of resolving the three reflections, the procedure are detailed in par. 8.2.4.
7.2.8
Penandaan
7.2.8
Marking.

Setiap Search Unit harus ditandai dengan jelas untuk mengindikasikan frekuensi dari tranduser, sudut nominal refleksi, serta index point yang digunakan

Each search unit shall be marked to clearly indicate the frequency of the transducer, nominal angle of refraction, and index point.
7.2.9
Jarak Ujung Probe
7.2.9
Edge Distance.

Dimensi dari Search Unit harus diatur jaraknya dari sudut leading dari Search Unit sampai index point tidak boleh melebihi 1 in. [25 mm].

The dimensions of the search unit shall be such that the distance from the leading edge of the search unit to the index point shall not exceed 1 in. [25 mm].




7.3
Standar Acuan
7.3
Reference Standards.
7.3.1
Standard IIW. The International Institute of Welding (IIW) UT reference block, sesuai  Gambar 6.22, sebagai standard yang dipakai untuk kalibrasi jarak dan sensitivitas. Block portable lain boleh digunakan, sebagai level acuan dari kombinasi sensitifitas instrument / Search Unit yang harus di sesuaikan dengan nilai yang sama pada IIW Block (Lihat Annex H dari AWS D 1.1, sebagai contoh).
7.3.1
IIW Standard. The International Institute of Welding (IIW) UT reference block, shown in Figure 6.22, shall be the standard used for both distance and sensitivity calibration. Other portable blocks may be used, provided the reference level sensitivity for instrument / search unit combination shall be adjusted to be the equivalent of that achieved with the IIW Block (see Annex H of AWS D 1.1, for examples).
7.3.2
Blok Test Acuan penguji Resolusi RC seperti pada Gambar 6.23.
7.3.2
The RC resolution reference test block shown in Figure 6.23.
7.3.3
Blok Acuan untuk jarak dan sensitivitas sesuai Gambar 6.26 dan 6.21
7.3.3
The  Distance and Sensitivity Reference Block shown in Figure 6.26 and 6.21




8
Instruksi – instruksi
8
Instructions
8.1
Prosedur Kalibrasi dan Kualifikasi   Peralatan
8.1
Equipment Calibration and Qualifi-cation Procedure
8.1.1
Linearitas Horizontal.
8.1.1
Horizontal Linearity.

Linearitas Horizontal dari instrumen penguji harus di rekualifikasi setelah setiap 40 jam penggunaan instrumen dalam setiap range jarak pada saat instrumen itu akan digunakan.
Prosedur Kualifikasi (lihat Annex H AWS D 1.1, untuk  metode alternative)  harus seperti ini :

Catatan:
Karena prosedur kualifikasi ini menggunakan straightbeam search unit yang menghasilkan gelombang longitudinal dengan sound velocity hampir dua kali gelombang shear, diperlukan untuk  menggandakan range jarak gelombang shear untuk aplikasi dari prosedur ini.

Contoh: Penggunaan 10 in. [250 mm] screen kalibrasi dalam gelombang shear akan meminta 20 in. [500 mm] kalibrasi screen untuk kualifikasi prosedur ini menggunakan straight-beam.
Prosedur berikut harus digunakan untuk kalibrasi instrument (lihat Annex H, H3, untuk metode alternatif):
(1)      Probe Normal memenuhi syarat par. 7.2.1 harus ditempatkan  pada IIW atau DS block Posisi G, T, atau  U (lih. Figure 6.26) sedemikian rupa  sehingga mendapatkan lima refleksi balik (pulsa) pada range kualifikasi yang telah disertifi-kasikan Figure 6.26).
(2)      Refleksi balik pertama dan kelima harus disesuaikan ke lokasi yang tepat dengan menggunakan kalibrasi jarak dan penyesuaian kelambatan nol.
(3)      Setiap indikasi harus disesuaikan terhadap level referensi dengan gain atau kontrol atenuasi untuk pengamatan lokasi horizontal
(4) Setiap defleksi lokasi pada sisa intermediate harus di disesuaikan dengan 2% dari lebar sreen.

The horizontal linearity of the test instrument shall be requalified after each 40 hours of instrument use in each of the distance ranges that the instrument will be used.
The qualification procedure (see Annex H AWS D 1.1, for alternative method)  shall be as follows :

Note:
Since this qualification procedure is performed with a straightbeam search unit which produces longitudinal waves with a sound velocity of almost double that of shear waves, it is necessary to double the shear wave distance ranges to be used in applying this procedure.

Example: The use of a 10 in. [250 mm] screen calibration in shear wave would require a 20 in. [500 mm] screen calibration for this qualification procedure using straight-beam.

The following procedure shall be used for instrument qualification (see Annex H, H3, for alternative method):
(1)      A straight-beam search unit meeting the requirements of par. 7.2.1 shall be coupled to the IIW or DS block in Position G, T, or U (see Figure 6.26) as necessary to attain five back reflections in the qualification range being certified (see Figure 6.26).

(2)      The first and fifth back reflections shall be adjusted to their proper locations with use of the distance calibration and zero delay adjustments.
(3)      Each indication shall be adjusted to reference level with the gain or attenuation control for horizontal location examination.
(4)      Each intermediate trace deflection location shall be correct within 2% of the screen width.
8.1.2
Gain Control (dB Accuracy ).
8.1.2
Gain Control (dB Accuracy ).

Gain kontrol instrumen (attenuator) harus dicek untuk kebenaran kalibrasinya setiap interval 2 (dua) bulan.
Prosedur qualifikasi harus dilaksanakan seperti berikut :

Note :
Untuk supaya menghasilkan keakurasian pembacaan ketinggian indikasi yang diinginkan (± 1% ), display vertikal harus di atur secara bertahap interval pada 2% , atau 2.5% untuk instrument dengan bacaan digital amplitude, pada ketinggian horizontal tengah screen.
Cara lain bisa digunakan untuk kalibrasi gain kontrol (attenuator) qualifikasi jika hasilnya minimum sama dengan prosedur ini.
Tahapan ini harus diletakkan pada display antara 60% dan 100% dari tinggi screen.
Ini bisa dilaksanakan dengan menggunakan selembar lapisan transparan pada screen. Jika lapisan transparan ini di aplikasikan sebagai bagian permanent dari UT unit, harus diperhatikan supaya lapisan tidak mengganggu display sewaktu test.

The instrument’s gain control (attenuator) shall be checked for correct calibration at 2 (two) month intervals.
The qualification procedure shall be performed as follows :

Note:
In order to attain the required accuracy (± 1% ) in reading the indication height, the display shall be graduated vertically at 2% intervals, or 2.5% for instruments with digital amplitude readout, at horizontal mid-screen height.

Alternative methods may be used for calibrated gain control (attenuator) qualification if proven at least equivalent with this procedure.
These graduations shall be placed on the display between 60% and 100% of screen height.
This may be accomplished with use of a graduated transparent screen overlay. If this overlay is applied as a permanent part of the UT unit, care should be taken that the overlay does not obscure normal testing displays.

(1)   Probe normal sesuai requirement dari par. 7.2.1 harus di letakkan  ke Blok kalibrasi DS  terlihat pada  Figure 6.23  dan pada posisi “ T “, Figure 6.26.
(2)   Kalibrasi jarak harus diatur sehingga refleksi indikasi back wall yang pertama dari 2 in. [50 mm]  ( yang kemudian dinamakan indikasi ) ada ditengah screen horizontal.
(3)   Gain atau attenuator kontrol yang terkalibrasi harus diatur sehingga indikasi tepat atau sedikit diatas 40% screen height.
(4)   Search unit harus digeser kearah posisi U, lihat Figure 6.26, sampai indikasi tepat pada 40% screen height.
(5)   Amplitude sound harus dinaikkan 6 dB dengan gain atau attenuator kontrol yang terkalibrasi. Indikasi level secara teori seharusnya tepat pada 80% screen height.

(1)   A straight-beam search unit meeting the requirements of par. 7.2.1 shall be coupled to the DS block kalibrasi shown in Figure 6.23 and position “ T ”, Figure 6.26.
(2)   The distance calibration shall be adjusted so that the first 2 in. [50 mm] back reflection indication (hereafter called the indication) is at horizontal mid-screen.

(3)   The calibrated gain or attenuation control shall be adjusted so that the indication is exactly at or slightly above 40% screen height.
(4)   The search unit shall be moved toward position U, see Figure 6.26, until the indication is at exactly 40% screen height.
(5)   (The sound amplitude shall be increased 6 dB with the calibrated gain or attenuation control. The indication level theoretically should be exactly at 80% screen height.

(6)   Pembacaan dB harus di record dengan “ a “ dan actual % screen height dengan “ b “ dari step 5 pada laporan certifikasi (Annex M, Form M-8), Line 1.
(7)   Search unit  harus digerakkan lebih mengarah ke posisi U, Figure 6.26, sampai indikasi tepat pada 40% screen height.
(8)   Step 5 harus diulangi.
(9)   Step 6 harus diulangi; kecuali informasi seharusnya di aplikasikan ke line berikutnya pada Annex M, Form M-8.
(10)  Step 7, 8, dan 9 harus diulangi secara urut sampai seluruh range  dari gain kontrol (attenuator) mencapai  (60 dB minimum).

(11)   Informasi dari row “ a “ dan “ b “ harus di aplikasikan ke rumus par. 8.1.3  untuk menghitung dB yang betul.
(12)  DB yang betul dari step 11 diisikan ke row “ c “.
(13)  Nilai row “ c “ harus dikurangi dari nilai row “ a “  dan perbedaan di row “ d “, dB error harus diisi. 

Note : Nilai ini bisa positive maupun negative . Contoh applikasi ada di Annex M. Forms M-8, M-9, and M-10 .

(14)  Informasi harus di tabulasikan dalam sebuah Form , termasuk minimum persamaan informasi seperti yang ditampakkan Form D-8, dan unit di evaluasi sesuai dengan instruksi yang ditunjukkan  pada form ini.
(15)  Form M-9 menyajikan cara relative lebih sederhana untuk mengevaluasi data dari item (14). Instruksi untuk evaluasi ini diberikan pada (16) sampai (18).
(16)  DB informasi dari row “ e “(Form M-8) harus diisikan vertikal dan pembacaan dB dari Row “a” (Form M-8) secara horizontal seperti X and Y coordinates untuk ploting dB curve pada Form M-9.
(17)Panjang horizontal terpanjang, seperti dipresentasikan oleh pembacaan perbedaan dB, yang di tentukan dalam sebuah segi panjang dalam tinggi 2 dB, menyatakan range dB dimana equipment memenuhi syarat2 Code.
(18)  Minimum range dB yang diizinkan adalah 60 dB.
(19)Equipment yang tidak memenuhi syarat mungkin bisa digunakan, asalkan faktor koreksi dicatat dan digunakan untuk evaluasi cacat menyimpang range linearity instrument yang acceptable, atau evaluasi test lasan dan  cacat dipertahankan dalam range vertikal linearity dari equipment. 


Note : DB error gambar ( Row “ d “ ) bisa digunakan untuk sebagai faktor koreksi gambar.


(6)   The dB reading shall be recorded under “a” and actual % screen height under “b” from step 5 on the certification report (Annex M, Form M-8), Line 1.
(7)   The search unit shall be moved further toward position U, Figure 6.26, until the indication is at exactly 40% screen height.
(8)   Step 5 shall be repeated.
(9)   Step 6 shall be repeated; except, information should be applied to the next consecutive line on Annex M, Form M-8.
(10)  Steps 7, 8, and 9 shall be repeated consecutively until the full range of the gain control (attenuator) is reached (60 dB minimum).
(11)  The information from Rows “a” and “b” shall be applied to equation par. 8.1.3  to calculate the corrected dB.
(12)  Corrected dB from step 11 to Row  “c ” shall be applied.
(13)  Row “c ” value shall be subtracted from Row “a ” value and the difference in Row “d,” dB error shall be applied.
Note: These values may be either positive or negative and so noted. Examples of Application of Forms M-8, M-9, and M-10 are found in Annex M.
(14)  Information shall be tabulated on a form, including minimum equivalent information as displayed on Form D-8, and the unit evaluated in conformance with instructions shown on that form.
(15)  Form M-9 provides a relatively simple means of evaluating data from item (14). Instructions for this evaluation are given in (16) through (18).
(16)  The dB information from Row “e” (Form M-8) shall be applied vertically and dB reading from Row “a” (Form M-8) horizontally as X and Y coordinates for plotting a dB curve on Form M-9.
(17)  The longest horizontal length, as represented by the dB reading difference, which can be inscribed in a rectangle representing 2 dB in height, denotes the dB range in which the equipment meets the code requirements.
(18)  The minimum allowable range is 60 dB.
(19)  Equipment that does not meet this minimum requirement may be used, provided correction factors are developed and used for discontinuity evaluation outside the instrument acceptable linearity range, or the weld testing and discontinuity evaluation is kept within the acceptable vertical linearity range of the equipment. 
Note: The dB error figures (Row “d”) may be used as correction factor figures.

8.1.3
Rumus Decibel.
8.1.3
Decibel Equation.

Rumus berikut harus digunakan untuk kalkulasi decibel :
Seperti dihubungkan dengan Annex M, Form M-8
dB1 = Row “a”
dB2 = Row “c”
 %1  = Row “b”
 %2  = Defined on Form M-8




The following equation shall be used to calculate decibels:
As related to Annex M, Form M-8

dB1 = Row “a”
dB2 = Row “c”
 %1  = Row “b”
 %2  = Defined on Form M-8

            
8.1.4
Procedure Refleksi Internal.
8.1.4
Internal Reflections Procedure.

Maximum internal reflections dari masing2 unit  harus di ferifikasi maximum waktu interval 40 jam selama instrument digunakan.
(1)   Kalibrasi equipment sesuai dengan par. 8.5.
(2)   Pindahkan search unit dari blok kalibrasi tanpa merubah pengatur equipment apapun.
(3)   Naikkan gain kontrol yang terkalibrasi atau attenuator 20 dB lebih sensitive dari reference level.

(4)   Area screen diluar 1/2 in. [12 mm] sound path dan diatas tinggi reference level harus bebas dari indikasi apapun.

Maximum internal reflections from each search unit shall be verified at a maximum time interval of 40 hours of instrument use in
(1)   Calibrate the equipment in conformance with par. 8.5.
(2)   Remove the search unit from the calibration block kalibrasi without changing any other equipment adjustments.
(3)   Increase the calibrated gain or attenuation 20 dB more sensitive than reference level.
(4)   The screen area beyond 1/2 in. [12 mm] sound path and above reference level height shall be free of any indication.

8.2
Kalibration dari Search Units.
8.2
Calibration of Search Units.
8.2.1
Kalibrasi dari Search Unit probe sudut
8.2.1
Calibration of Angle-Beam Search Units.

Dengan menggunakan Blok Kalibrasi yang diaprove, setiap angle - beam search unit harus dicek setiap delapan jam setelah digunakan, bahwa angle - beam search unit masih betul, dan bahwa sudut probe masih dalam toleransi yang diizinkan plus atau minus 2°.
Search unit yang tidak memenuhi syarat harus dibetulkan atau dibuang.

With the use of an approved calibration block, each angle - beam search unit shall be checked after each eight hours of use to determine that the contact face is flat, that the sound entry point is correct, and that the beam angle is within the allowed plus or minus 2° tolerance .
The Search units which do not meet these requirements shall be corrected or replaced.
8.2.2
Titik Index.
8.2.2
Index point.

Index point dari transducer harus ditentukan atau dicek dengan procedure sebagai berikut :
(1)   Transducer harus di set pada posisi D pada the IIW block.
(2)   Transducer harus di gerakkan sampai signal dari radius menjadi maximum. Titik pada transducer dimana segaris dengan garis radius pada calibration block adalah titik masuknya Ultrasonic  (Index point ).

The transducer sound entry point shall be located or checked by the following procedure:
(1)      The transducer shall be set in position D on the IIW block.
(2)      The transducer shall be moved until the signal from the radius is maximized. The point on the transducer which aligns with the radius line on the calibration block is the point of sound entry .

8.2.3
Sudut
8.2.3
Angles.

Search unit harus menghasilkan sound beam dalam material yang ditest dalam range plus atau minus 2° dari salah satu dari sudut yang betul sebagai : 70°, 60°, atau 45°.
Sound path transducer harus dicek atau ditentukan dengan  procedure berikut :
(1)   Transducer harus di set pada posisi B pada IIW block  untuk sudut  40° sampai 60°, atau pada posisi C pada IIW block untuk sudut 60° sampai 70° (lih. Figure 6.26).

(2)   Untuk sudut yang ditentukan, transducer harus di gerakkan ke muka dan ke belakang diatas garis – garis sudut yang diperkirakan sampai signal dari radius menjadi maximum.
Index point pada transducer harus di bandingkan dengan sudut yang tertera pada calibration block (tolerance ± 2°)

The search unit shall produce a sound beam in the material being tested within plus or minus 2° of one of the following proper angles: 70°, 60°, or 45°.
The transducer sound-path angle shall be checked or determined by one of the following procedures:
(1) The transducer shall be set in position B on IIW block for angles 40° through 60°, or in position C on IIW block for angles 60° through 70° (see Figure 6.26).
(2) For the selected angle, the transducer shall be moved back and forth over the line indicative of the transducer angle until the signal from the radius is maximized.

The sound entry point on the transducer shall be compared with the angle mark on the calibration block (tolerance ± 2°)

Jika UT di aplikasikan pada stainless steel material  , sudut untuk kalkulasi harus sebagai berikut :

When UT is applied on stainless steel material, the angle for calculating shall be as follows :

Sudut di
 Carbon and Alloy steel
70 °
60 °
45 °
Sudut di
Stainless steel

65 °
57 °
43 °

Angle in
Carbon and Alloy steel
70 °
60 °
45 °
Angle in
Stainless steel

65 °
57 °
43 °

8.2.4
Persyaratan untuk resolusi dari probe sudut .
8.2.4
Angle Beam Resolution Require-ments.

Kombinasi search unit dan instruments harus bisa membedakan tiga lubang dalam RC resolution reference test block kalibrasi ditunjukkan pada Figure 6.23.
Posisi search unit seperti berikut :
(1)   Transducer harus di set pada resolution block RC posisi Q for 70° angle, posisi R untuk 60° angle, atau posisi S untuk 45° angle.
(2)   Transducer dan instrument harus bisa membedakan tiga lubang, paling tidak membedakan ujung indikasi dari tiga lubang.

Resolution harus dievaluasi dengan instrument control diset pada normal setting dan dengan indikasi dari lubang dibawa ke tengah screen height. Resolution harus cukup membedakan ujung indikasi dari tiga lubang. Menggunakan RC resolution reference block kalibrasi untuk kalibrasi harus dilarang. Setiap kombinasi dari instrument search unit (shoe dan transducer) harus dicek sebelum pertama kali dipergu-nakan. Equipment verification ini harus dikerjakan pertama kali dengan setiap search unit adan UT combinasi.

The combination of search unit and instrument shall resolve three holes in the
RC resolution reference test block shown in Figure 6.23.
The search unit position as follows :
(1)   The transducer shall be set on resolution block RC position Q for 70° angle, position R for 60° angle, or position S for 45° angle.
(2)   Transducer and instrument shall resolve the three test holes, at least to the extent of distinguishing the peaks of the indications from the three holes.
The resolution shall be evaluated with the instrument controls set at normal test settings and with indications from the holes brought to midscreen height. Resolution shall be sufficient to distinguish at least the peaks of indications from the three holes. Use of the RC resolution reference block for calibration shall be prohibited. Each combination of instrument search unit (shoe and transducer) shall be checked prior to its initial use. This equipment verification shall be done initially with each search unit and UT unit combination.

Verifikasi tidak perlu dikerjakan lagi asalkan dokumentasi di simpan yang merekord seperti dibawah ini :

(1)       Merk alat UT , model dan nomor seri.
(2)       Pabrik pembuat search unit , tipe, ukuran, sudut, dan nomor seri.
(3)       Tanggal verivikasi dan nama yang melakukan.

The verification need not be done again provided documentation is maintained that records the following items:
(1)      UT machine’s make, model and serial number
(2)      Search unit’s manufacturer, type, size, angle, and serial number
(3) Date of verification and technician’s name
8.2.5
Kalibrasi dari Search Units Probe Normal (Longitudinal Mode ).
8.2.5
Calibration of Straight-Beam Search Units (Longitudinal Mode ).

Kalibrasi Jarak. Lihat See Annex H, H1 untuk  alternative method.
(1)      Transducer harus diset pada posisi G pada  the IIW block.
(2)      Instrument  harus diatur untuk bisa meng-hasilkan indikasi pada 1 in. [25 mm pada a metric block], 2 in. [50 mm pada a metric block], 3 in. [75 mm pada a metric block], 4 in. [100 mm pada a metric block], etc., pada display.

Distance Calibration. See Annex H, H1 for alternative method.
(1) The transducer shall be set in position G on the IIW block.
(2) The instrument shall be adjusted to produce indications at 1 in. [25 mm on a metric block], 2 in. [50 mm on a metric block], 3 in. [75 mm on a metric block], 4 in. [100 mm on a metric block], etc., on the display.

Amplitude. Lihat Annex H, H1.2 untuk alternative method.
(1)      Transducer harus diset pada posisi Gain pada IIW block.
(2)      Gain harus diatur sampai indikasi maximum dari back reflection yang pertama mencapai 50 sampai 75% screen height.

Amplitude. See Annex H, H1.2 for alternative method.
(1)      The transducer shall be set in position G on the IIW block.
(2) The gain shall be adjusted until the maximized indication from first back reflection attains 50 to 75% screen height.
8.2.6
Persyaratan Resolution Probe Normal
8.2.6
Straight-Beam Resolution Requi-rement

(1)       Transducer harus diset pada posisi F pada IIW block.
(2)       Transducer dan instrument harus bisa membedakan semua tiga jarak.

(1)       The transducer shall be set in position F on the IIW block.
(2) Transducer and instrument shall resolve all three distances.
8.2.7
Kualifikasi Linearitas Horizontal
8.2.7
Horizontal Linearity Qualification.

Procedure Qualifikasi harus seperti par. 8.1.1.

Qualification procedure shall be per par. 8.1.1.
8.2.8
Kualifikasi Gain Control (Attenuation).
8.2.8
Gain Control (Attenuation) Qualifi-cation.

Procedure Qualifikasi harus seperti par. 8.1.2.

The qualification procedure shall be in conformance with par. 8.1.2.




8.3
Kalibration untuk  Pengujian
8.3
Calibration for Testing
8.3.1
Posisi Kontrol Reject.
8.3.1
Position of Reject Control.

Semua kalibrasi dan test harus dibuat dengan reject (clipping or suppression) control pada posisi mati. Menggunakan reject (clipping or suppression) control bisa mengubah amplitude linearity dari the instrumen dan hasil test yang tidak valid.

All calibrations and tests shall be made with the reject (clipping or suppression) control turned off. Use of the reject (clipping or suppression) control may alter the amplitude linearity of the instrument and invalidate test results.
8.3.2
Reflector yang dilarang.
8.3.2
Prohibited Reflectors.

Menggunakan reflector pojok (corner) untuk kalibrasi tidak diperbolehkan.

The use of a “corner” reflector for calibration purposes shall be prohibited.
8.3.3
Teknik.
8.3.3
Technique.

Kalibrasi untuk sensitivity dan horizontal Sweep ( jarak ) harus dibuat oleh UT operator sesaat sebelum dan pada lokasi dimana testing lasan dimulai.

Calibration for sensitivity and horizontal sweep (distance) shall be made by the UT operator just prior to and at the location of testing of each weld.
8.3.4
Kalibrasi Ulang.
8.3.4
Recalibration.

Re kalibrasi harus dibuat setelah penggantian operator, setiap 30 menit waktu maximum interval, atau aliran listrik terganggu dengan meliputi :
(1)      Penggantian transducer
(2)      Penggantian baterei
(3)      Penggantian kabel listrik
(4)      Penggantian kabel probe
(5)      Power outage ( kegalalan )

Recalibration shall be made after a change of operators, each 30 minute maximum time interval, or when the electrical circuitry is disturbed in any way which includes the following:
(1)   Transducer change
(2)   Battery change
(3)   Electrical outlet change
(4)   Coaxial cable change
(5)   Power outage (failure)




8.4
Pengujian Probe Normal  untuk  Base Metal.
8.4
Straight-Beam Testing of Base Metal.
8.4.1
Kalibrasi untuk straight - beam testing dari base metal harus dibuat dengan search unit dari permukaan A dari base metal dan dikerjakan sebagai berikut :
8.4.1
Calibration for straight-beam testing of base metal shall be made with the search unit applied to Face A of the base metal and performed as follows:
8.4.2
Sweep Horizontal.
8.4.2
Horizontal Sweep.

Horizontal sweep harus diatur untuk jarak kalibrasi dengan setara minimum dua kalibrasi ketebalan plate pada display.

The horizontal sweep shall be adjusted for distance calibration to present the equivalent of at least two plate thicknesses on the display.
8.5
Kalibrasi untukAngle-Beam Testing.
8.5
Calibration for Angle-Beam Testing.

Kalibrasi untuk angle – beam harus dibuat seperti berikut (see Annex H, H2.4 for alternative method).

Calibration for angle-beam testing shall be performed as follows (see Annex H, H2.4 for alternative method).
8.5.1
Horizontal Sweep.
8.5.1
Horizontal Sweep.

Horizontal sweep harus diatur untuk mewakili actual jarak sound-path dengan menggunakan IIW block, dengan transducer diset pada posisi dengan pada IIW block. Jarak kalibrasi harus dibuat menggunakan salah satu 5 in. [125 mm] scale atau 10 in. [250 mm] scale pada display, yang mana yang lebih pantas. Jika joint konfigurasi atau tebal tidak memungkinkan untuk pengujian secara full dari lasan dari masing – masing seting , kalibrasi jarak harus di buat menggunakan skala 15 in. atau 20 in. [400 mm atau 500 mm] sesuai kebutuhan.
Note : Lokasi horizontal dari indikasi screen didasarkan pada lokasi dimana sisi kiri dari deflection memotong garis horizontal .

The horizontal sweep shall be adjusted to represent the actual sound-path distance by using the IIW block, with transducer shall be set in position D on the IIW block. The distance calibration shall be made using either the 5 in. [125 mm] scale or 10 in. [250 mm] scale on the display, whichever is appropriate. If, however, the joint configuration or thickness prevents full examination of the weld at either of these settings, the distance calibration shall be made using 15 in. or 20 in. [400 mm or 500 mm] scale as required.

Note: The horizontal location of all screen indications is based on the location at which the left side of the trace deflection breaks the horizontal base line.
8.5.2
Zero Reference Level.
8.5.2
Zero Reference Level.

Zero reference level sensitivity yang digunakan untuk evaluasi cacat (“b” on the Ultrasonic Test AWS, terlampir) harus dihasilkan dengan mengatur gain kontrol kalibrasi (attenuator) detector cacat, memenuhi persyaratan dari par. 8.1.2, sehingga maximum horizontal trace deflection (diatur ke horizontal ketinggian reference line dengan gain kontrol yang terkalibrasi [attenuator]) terlihat pada display.
Amplitude atau Sensitivity Calibration Procedure, seperti berikut :
Transducer harus diset pada posisi A pada IIW block (sembarang sudut). Sinyal yang maximum dari lubang 0.060 in. [1.59 mm] kemudian harus diatur untuk menghasilkan ketinggian indikasi horizontal reference-line (lihat Annex H, H2.4 untuk alternative method). Maximum pembacaan decibel yang dihasilkan harus digunakan sebagai pembacaan  “Reference Level, b” pada Test Report sheet (Ultrasonic Test AWS , attached).


The zero reference level sensitivity used for discontinuity evaluation (“b” on the Ultrasonic Test AWS, attached) shall be attained by adjusting the calibrated gain control (attenuator) of the discontinuity detector, meeting the requirements of par. 8.1.2, so that a maximized horizontal trace deflection (adjusted to horizontal reference line height with calibrated gain control [attenuator]) results on the display.
Amplitude or Sensitivity Calibration Procedure, as follow :
The transducer shall be set in position A on the IIW block (any angle). The maximized signal shall then be adjusted from the 0.060 in. [1.59 mm] hole to attain a horizontal reference-line height indication (see Annex H, H2.4 for alternative method). The maximum decibel reading obtained shall be used as the “Reference Level, b” reading on the Test Report sheet (Ultrasonic Test AWS , attached).

8.6
Procedur Pengujian
8.6
Testing Procedures
8.6.1
“X” Line.
8.6.1
“X” Line.

Sebuah garis “ X “ untuk lokasi cacat harus ditandai pada permukaan test dari lasan sejajar dengan arah dari garis tengah lasan. Jarak lokasi tegak lurus dengan garis tengah lasan harus didasarkan  pada dimensi gambar pada detail drawing dan biasanya jatuh pada centerline  dari lasan butt joint, dan selalu jatuh pada permukaan terdekat connecting member dari  Testing dan lasan corner joint (dibaliknya Permukaan C).

An “X” line for discontinuity location shall be marked on the test face of the weldment in a direction parallel to the weld axis. The location distance perpendicular to the weld axis shall be based on the dimensional figures on the detail drawing and usually falls on the centerline of the butt joint welds, and always falls on the near face of the connecting member of T and corner joint welds (the face opposite Face C).
8.6.2
“Y” Line.
8.6.2
“Y” Line.

Sebuah garis “ Y “ bersama nomor identifikasi lasan harus jelas ditandai pada base metal dekat dengan lasan yang ditest dengan UT. Marking ini digu-nakan untuk maksud sebagai berikut :
(1)      Identifiksai lasan
(2)      Identifikasi dari permukaan A
(3)      Jarak pengukuran dan arah (+ atau –) dari garis “X”
(4)      Pengukuran lokasi dari ujung dan akhir lasan

A “Y” accompanied with a weld identification number shall be clearly marked on the base metal adjacent to the weld that is subject to UT. This marking is used for the following purposes:
(1)   Weld identification
(2)   Identification of Face A
(3)   Distance measurements and direction (+ or –) from the “X” line
(4)   Location measurement from weld ends or edges
8.6.3
Kebersihan.
8.6.3
Cleanliness.

Semua permukaan dimana search unit akan digunakan harus bebas dari spatter lasan, debu, grease, oil (selain yang digunakan sebagai couplant), cat, dan kerak yang mengelupas dan harus mempunyai contour yang memungkinkan menempel bagus.

All surfaces to which a search unit is applied shall be free of weld spatter, dirt, grease, oil (other than that used as a couplant), paint, and loose scale and shall have a contour allowing intimate coupling.

8.6.4
Couplants.
8.6.4
Couplants.

Couplant material harus digunakan diantara search unit dan material yng ditest. Couplant harus dari salah satu glycerin atau campuran yang merata dari cellulose gum dan air. Pelumas bisa ditambahkan jika diperlukan.

Oli mesin yang encer bisa digunakan untuk couplant pada block kalibrasi.

A couplant material shall be used between the search unit and the test material. The couplant shall be either glycerin or cellulose gum and water mixture of a suitable consistency. A wetting agent may be added if needed.
Light machine oil may be used for couplant on calibration blocks.
8.6.5
Maksud dari pengujian.
8.6.5
Extent of Testing.

Seluruh base metal dimana harus dilalui oleh ultarasound untuk pengujian lasan harus diuji kemungkinan laminasi menggunakan straight - beam search unit sesuai persyaratan dari par. 7.2.1 dan di kalibrasi sesuai dengan par. 8.4. Jika ada area dari base metal menunjukkan back reflection hilang total atau ada indikasi yang sama dengan atau lebih besar dari ketinggian original back reflection yang terletak pada posisi dimana akan mengganggu procedure scaning normal  lasan, maka ukurannya, lokasinya, dan kedalamannya dari permukaan A harus ditentukan dan dilaporkan pada UT report, alternative procedure scaning lasan harus digunakan.

The entire base metal through which ultrasound must travel to test the weld shall be tested for laminar reflectors using a straight-beam search unit conforming to the requirements of par. 7.2.1 and calibrated in conformance with par. 8.4. If any area of base metal exhibits total loss of back reflection or an indication equal to or greater than the original back reflection height is located in a position that will interfere with the normal weld scanning procedure, its size, location, and depth from the A face shall be determined and reported on the UT report, and an alternate weld scanning procedure shall be used.
8.6.6
Kesulitan access.
8.6.6
Inaccessibility.

Jika bagian dari lasan susah di access untuk pengujian sesuai dengan persyaratan dari Table 6.7, karena adanya record laminasi sesuai dengan par. 8.6.5, pengujian harus dilaksanakan menggunakan satu atau lebih procedure alternative  sesuai kebutuhan untuk mengahasilkan seluruh lasan tercover:
(1)      Permukaan lasan harus digerinda rata. Jika finishing permukaan diperlukan , nilai kekasaran ( lihat ASME B46.1) tidak boleh lebih dari 250 microinches [6.3 micrometers]. Nilai Surface finish melebihi125 micro inche [3.2 micrometer] sampai 250 microinches harus dikerjakan sejajar dengan arah dari primary stress. Nilai Surface finish  125 microinches atau kurang bisa dikerjakan pada sembarang arah.
(2)      Pengujian dari permukaan A dan B harus dilakukan.
(3)      Sudut Search unit lain harus digunakan.

If part of a weld is inaccessible to testing in conformance with the requirements of Table 6.7, due to laminar content recorded in conformance with par. 8.6.5, the testing shall be conducted using one or more of the following alternative procedures as necessary to attain full weld coverage:
(1)      Weld surface(s) shall be ground flush . Where surface finishing is required, roughness values (see ASME B46.1) shall not exceed 250 microinches [6.3 micrometers]. Surfaces finished to values of over 125 microinches [3.2 micrometers] through 250 microinches shall be finished parallel to the direction of primary stress. Surfaces finished to values of 125 microinches or less may be finished in any direction.


(2)      Testing from Faces A and B shall be performed.
(3)      Other search unit angles shall be used.




8.7
Pengujian Lasan.
8.7
Testing of Welds.
8.7.1
Lasan harus diuji dengan menggunakan probe sudut sesuai dengan persyaratan dari par. 7.2.3 ~ 7.2.9 dengan instrument yang terkalibrasi sesuai dengan par. 8.2.1 menggunakan sudut seperti ditunjukkan dalam Table 6.7. Setelah kalibrasi dan selama pengujian, pengaturan kontrol  instrument yang diperbolehkan hanya pengaturan sensitivity dengan gain kontrol yang terkalibrasi (attenuator).
8.7.1
Welds shall be tested using an angle beam search unit conforming to the requirements of par. 7.2.3 ~ 7.2.9 with the instrument calibrated in conformance with par. 8.2.1 using the angle as shown in Table 6.7. Following calibration and during testing, the only instrument adjustment allowed is the sensitivity level adjustment with the calibrated gain control (attenuator).
8.7.2
Reject (clipping or suppression) kontrol harus dimatikan. Sensitivity harus dinaikan dari reference level untuk scaning lasan sesuai dengan Table 6.2 atau  6.3, yang sesuai.
8.7.2
The reject (clipping or suppression) control shall be turned off. Sensitivity shall be increased from the reference level for weld scanning in conformance with Table 6.2 or 6.3, as applicable.




8.8
Scanning.
8.8
Scanning.

Sudut pengujian dan procedure scaning harus sesuai dengan yang ditunjukan dalam Table 6.7.
Scaning Patern (lihat Figure 6.24)

The testing angle and scanning procedure shall be in conformance with those shown in Table 6.7.
Scanning Patterns (See Figure 6.24)




8.8.1
Longitudinal Discontinuities :
8.8.1
Longitudinal Discontinuities :

(1)   Scaning Movement A. Sudur putar, a = 100.
(2)   Scaning Movement B. Scaning distance b harus  mengcover semua bagian dari lasan.

(3)   Scanning Movement C. Jarak maju c harus kira – kira setengah lebar transducer .

Note : pergerakan A, B dan C bisa dikombinasikan menjadi satu scaning patern.

(1)       Scanning Movement A. Rotation angle, a = 10°.
(2)       Scanning Movement B. Scanning distance b shall be such that the section of weld being tested is covered.
(3)       Scanning Movement C. Progression distance c shall be approximately one-half the transducer width.
Note: movements A, B, and C may be combined into one scanning pattern.

8.8.2
Transverse Discontinuities :
8.8.2
Transverse Discontinuities :

(1)      Lasan yang digerinda.
Scaning patern D  harus  digunakan jika lasan digerinda rata.
(2)      Lasan yang tidak digerinda. 
Scanning pattern E harus digunakan jika weld reinforcement tidak digerinda rata.
Scaning angle, e = 15° max.

(1)      Ground Welds. Scanning pattern D shall be used when welds are ground flush.

(2)      Unground Welds. Scanning pattern E shall be used when the weld reinforcement is not ground flush.

Scanning angle, e = 15° max.

Note: Scaning patern harus mengcover semua bagian lasan.

Note: The scanning pattern shall cover the full weld section.




8.9
Butt Joints.
8.9
Butt Joints.
8.9.1
Semua butt joint harus diuji dari setiap sisi dari lasan dari garis tengah lasan. Lasan corner dan T joint harus diuji pertama dari satu sisi dari garis tengah lasan saja.
8.9.1
All butt joint welds shall be  tested from each side of the weld axis. Corner and T-joint welds shall be primarily tested from one side of the weld axis only.
8.9.2
Semua lasan harus diuji menggunakan scaning patern yang cocok atau patern yang ditunjukkan di Figure 6.24 jika diperlukan untuk mendeteksi kedua – duanya cacat memanjang  dan melintang.
8.9.2
All welds shall be tested using the applicable scanning pattern or patterns shown in Figure 6.24 as necessary to detect both longitudinal and transverse discontinuities.
8.9.3
Jika diperlukan, minimum, semua lasan diuji dengan sound melewati seluruh volume lasan dan HAZ dalam dua arah yang berpotongan, jika memungkinkan untuk  dilaksanakan.
8.9.3
It is intended that, as a minimum, all welds be tested by passing sound through the entire volume of the weld and the HAZ in two crossing directions, wherever practical.
8.9.4
Indikasi Maximum .
Jika sebuah indikasi cacat kelihatan di layar, maximum indikasi yang dihasilkan dari cacat harus diatur untuk menghasilkan horizontal reference level trace deflection pada display. Pengaturan ini harus dibuat dengan  gain kontrol terkalibrasi (attenuator), dan pembacaan decibel instruments harus digunakan sebagai “Indication Level, a,” untuk kalkulasi “Indication Rating, d,” seperti ditunjukkan pada test report (Utrasonic  Test AWS, attached).
8.9.4
Maximum Indication.
When a discontinuity indication appears on the screen, the maximum attainable indication from the discontinuity shall be adjusted to produce a horizontal reference level trace deflection on the display. This adjustment shall be made with the calibrated gain control (attenuator), and the instrument reading in decibels shall be used as the “Indication Level, a,” for calculating the “Indication Rating, d,” as shown on the test report (Utrasonic  Test AWS, attached).
8.9.5
Attenuation Factor.
“Attenuation Factor, c,” pada test report harus dihasilkan dengan mengurangi 1 inch (25 mm) dari jarak sound-path dan mengalikan sisanya dengan 2. Factor ini harus dibulatkan ke nilai dB yang terdekat. Nilai pecahan kurang dari 1/2 dB harus dikurangi ke dB level yang lebih rendah dan nilai pecahan 1/2 dB atau lebih besar dinaikkan ke level yang lebih tinggi.
8.9.5
Attenuation Factor.
The “Attenuation Factor, c,” on the test report shall be attained by subtracting 1 in. [25 mm] from the sound-path distance and multiplying the remainder by 2. This factor shall be rounded out to the nearest dB value. Fractional values less than 1/2 dB shall be reduced to the lower dB level and those of 1/2 dB or greater increased to the higher level.
8.9.6
Indication Rating.
“Indication Rating, d,” pada Utrasonic  Test AWS, attached, menggambarkan perbedaan perhitungan decibel antara indication level dan reference level dengan koreksi untuk atenuasi seperti ditujukkan dibawah ini .

Instruments dengan  gain pada dB:
d =  a – b – c
8.9.6
Indication Rating.
The “Indication Rating, d,” in the Utrasonic  Test AWS, attached, represents the algebraic difference in decibels between the indication level and the reference level with correction for attenuation as indicated in the following expressions.
Instruments with gain in dB:
d =  a – b – c




9
Procedures Evaluasi  Ukuran Discontinuities.
9
Discontinuity Size Evaluation Procedures
9.1
Pengujian dengan Probe Normal (Longitudinal)
9.1
Straight-Beam (Longitudinal) Testing.
9.1.1
Ukuran dari cacat laminasi tidak selalu mudah ditentukan, terutama yang ukurannya lebih kecil dari ukuran transducer .
9.1.1
The size of lamellar discontinuities is not always easily determined, especially those that are smaller than the transducer size.
9.1.2
Jika cacat lebih besar dari transducer, kehilangan total dari back reflection akan terjadi dan 6 dB kehilangan amplitude  (50% screen height) dan pengukuran ke centerline dari transducer biasanya dipercaya untuk menentukan ujung cacat.
9.1.2
When the discontinuity is larger than the transducer, a full loss of back reflection will occur and a 6 dB loss of amplitude and measurement to the centerline of the transducer is usually reliable for determining discontinuity edges.
9.1.3
Walaupun begitu, ukuran kira – kira dari evaluasi reflector, yang lebih kecil dari transducer, harus dibuat dengan cara mulai dari arah luar cacat dengan instruments yang terkalibrasi menggerakkan transducer kearah area cacat sampai indikasi terlihat pada display. Pada point ujung dari search unit ini adalah  dianggap ujung dari cacat .
9.1.3
However, the approximate size evaluation of those reflectors, which are smaller than the transducer, shall be made by beginning outside of the discontinuity with equipment calibrated moving the transducer toward the area of discontinuity until an indication on the display begins to form. The leading edge of the search unit at this point is indicative of the edge of the discontinuity.

9.2
Pengujian dengan Probe Sudut (Shear).
9.2
Angle-Beam (Shear) Testing.
9.2.1
Procedure berikut ini harus digunakan untuk menentukan panjang dari indikasi yang mempunyai dB rating yang lebih serius dari indikasi Class D. Panjang dari indikasi harus ditentukan dengan mengukur jarak antara lokasi centerline transducer dimana amplitude indication rating turun (drop) 50% (6 dB) dibawah rating untuk clasifikasi cacat yang cocok.

Panjang ini harus direcord pada kolom “discontinuity length” pada test report.
Jika diperlukan karena memandang amplitude cacat , procedure ini harus diulangi untuk menentukan panjang cacat dari clas A, B dan Cacat .
9.2.1
The following procedure shall be used to determine lengths of indications which have dB ratings more serious than for a Class D indication. The length of such indication shall be determined by measuring the distance between the transducer centerline locations where the indication rating amplitude drops 50% (6 dB) below the rating for the applicable discontinuity classification. This length shall be recorded under “discontinuity length” on the test report.
Where warranted by discontinuity amplitude, this procedure shall be repeated to determine the length of Class A, B, and C discontinuities.
9.2.2
Indikasi dari cacat yang tetap ada pada display selama search unit digerakkan maju dan mundur dari cacat (scanning movement “b”) bisa jadi menunjukkan cacat planar dengan dimensi cenderung  kearah throat.
9.2.2
Indications of discontinuities that remain on the display as the search unit is moved towards and away from the discontinuity (scanning movement “b”) may be indicative of planar discontinuities with significant through-throat dimension.
9.2.3
Karena major reflecting surface dari cacat yang critical minimum orientasinya 20° (untuk   70° search unit) dan 45° (untuk  45° search unit) dari tegaklurus ke sound beam, maka amplitude evaluasi (dB rating) tidak memungkinkan disposisi secara tepat.
9.2.3
Since the major reflecting surface of the most critical discontinuities is oriented a minimum of 20° (for a 70° search unit) to 45° (for a 45° search unit) from perpendicular to the sound beam, amplitude evaluation (dB rating) does not allow reliable disposition.
9.2.4
Jika indikasi menunjukkan ciri – ciri planar ada pada scaning sensitivity, evaluasi lebih detail dari cacat dengan cara lain sangat diperlukan (seperti alternative UT tehnik, RT, grinding atau gouging untuk  visual inspection, etc.).
9.2.4
When indications exhibiting these planar characteristics are present at scanning sensitivity, a more detailed evaluation of the discontinuity by other means shall be required (e.g., alternate UT techniques, RT, grinding or gouging for visual inspection, etc.)




10
Kriteria Penerimaan.
10
Acceptance Criteria.
10.1
Pengujian lasan dengan Utrasonic adalah dievaluasi berdasarkan refleksi ultrasound dari cacat yang secara proporsional menimbulkan effect pada intregitas dari lasan .
10.1
Ultrasonically tested welds are evaluated on the basis of a discontinuity reflecting ultrasound in proportion to its effect on the integrity of the weld.
10.2
Setiap cacat lasan harus di terima (accepted) atau ditolak (rejected) berdasar pada indication ratingnya dan panjangnya, sesuai dengan Table 6.2 untuk  statically loaded structures atau Table 6.3 untuk  cyclically loaded structures, mana yang cocok.
10.2
Each weld discontinuity shall be accepted or rejected on the basis of its indication rating and its length, in conformance with
Table 6.2 for statically loaded structures or Table 6.3 for cyclically loaded structures, whichever is applicable.
10.3
Hanya cacat – cacat yang direject perlu direcord pada test report, kecuali untuk lasan yang disebut di kontrak sebagai “Fracture Critical”, rating acceptable dalam range 6 dB, inclusive, minimum rating rejectable harus direcord pada test report.
10.3
Only those discontinuities which are rejectable need be recorded on the test report, except that for welds designated in the contract documents as being “Fracture Critical”, acceptable ratings that are within 6 dB, inclusive, of the minimum rejectable rating shall be recorded on the test report.




10.4
Kriteria Penerimaan untuk Statically Loaded  Nontubular Connections.
10.4
Acceptance Criteria for Statically Loaded  Nontubular Connections.
10.4.1
Acceptance criteria untuk lasan yang dipersyaratkan  UT sebagai tambahan dari visual inspection harus memenuhi persyaratan dari Table 6.2
10.4.1
The acceptance criteria for welds subject to UT in addition to visual inspection shall meet the requirements of Table 6.2
10.4.2
Untuk lasan CJP web-to flange, acceptance  dari cacat yang terdeteksi dengan pergerakan scaning selain dari scaning patern ‘E’ (lihat par. 8.8 ) bisa berdasarkan pada ketebalan lasan sama dengan actual web thickness plus 1 in. [25 mm]. Cacat yang terdeteksi dengan scaning patern ‘E’ harus dievaluasi kepada criteria dari Table 6.2 untuk  actual web thickness.
10.4.2
For CJP web-to flange welds, acceptance of discontinuities detected by scanning movements other than scanning pattern ‘E’ (see par. 8.8) may be based on weld thickness equal to the actual web thickness plus 1 in. [25 mm]. Discontinuities detected by scanning pattern ‘E’ shall be evaluated to the criteria of Table 6.2 for the actual web thickness.
10.4.3
Jika lasan CJP web-to flange menanggung beban tarik (tensile stress) tegak lurus ke lasan, mereka harus sesuai dengan persyaratan di Table 6.2.
10.4.3
When CJP web-to-flange welds are subject to calculated tensile stress normal to the weld, they should be so designated on the design drawing and shall conform to the requirements of Table 6.2.




10.5
Kriteria Penerimaan untuk Cyclically Loaded Nontubular Connections.
10.5
Acceptance Criteria for Cyclically Loaded Nontubular Connections.
10.5.1
Acceptance criteria untuk lasan yang dipersyaratkan  UT sebagai tambahan dari visual inspection harus memenuhi persyaratan berikut :
(1)      Lasan yang menerima beban tensile dengan segala kondisi dari beban harus sesuai dengan persyaratan dari Table 6.3.
(2)      Lasan yang menerima beban tekan harus sesuai dengan persyaratan dari Table 6.2.
10.5.1
The acceptance criteria for welds subject to UT in addition to visual inspection shall meet the following requirements:
(1)      Welds subject to tensile stress under any condition of loading shall conform to the requirements of Table 6.3.
(2) Welds subject to compressive stress shall conform to the requirements of Table 6.2.
10.5.2
Lasan CJP web-to-flange harus sesuai dengan persyaratan dari Table 6.2.
10.5.2
CJP web-to-flange welds shall conform to the requirements of Table 6.2.
10.5.3
Acceptance  untuk cacat yang terdeteksi dengan gerakan scaning selain dari scaning patern ‘E’ (lihat par. 8.8) bisa berdasarkan pada ketebalan lasan sama dengan actual web thickness plus 1 in. [25 mm].
10.5.3
Acceptance for discontinuities detected by scanning movements other than scanning pattern ‘E’ (see par. 8.8) may be based on a weld thickness equal to the actual web thickness plus 1 in. [25 mm].
10.5.4
Cacat yang dideteksi dengan scaning patern ‘E’ harus dievaluasi dengan criteria par. 10.5 untuk aktual ketebalan web.
10.5.4
Discontinuities detected by scanning pattern ‘E’ shall be evaluated to the criteria of par. 10.5 for the actual web thickness.
10.5.5
Jika lasan CJP web-to flange menerima beban tarik (tensile stress) tegak lurus ke lasan, mereka harus dijelaskan di drawing design dan harus sesuai dengan persyaratan di Table 6.3.
10.5.5
When such web-toflange welds are subject to calculated tensile stress normal to the weld, they shall be so designated on design drawings and shall conform to the requirements of Table 6.3.




11
Identification pada Area yang di Reject.
11
Identification of Rejected Area.

Setiap cacat yang direject harus ditandai pada lasan dengan tanda langsung diatas cacat untuk keseluruhan panjangnya. Kedalaman dari surface dan indication rating harus di tulis dekat base metal.

Each rejectable discontinuity shall be indicated on the weld by a mark directly over the discontinuity for its entire length. The depth from the surface and indication rating shall be noted on nearby base metal.




12
Perbaikan.
12
Repair.

Lasan yang ditemukan tidak diterima dengan UT harus direpair dengan metode yang diizinkan oleh par. 5.26 dari Code ini. Area yang direpair harus di uji ulang dengan ultrasonic  dengan hasil yang ditabulasikan kedalam form asli atau form report tambahan.

Welds found unacceptable by UT shall be repaired by methods allowed by 5.26 of this code. Repaired  areas shall be retested ultrasonically with results tabulated on the original form (if available) or additional report forms.





13
Laporan Pengujian Ulang.
13
Retest Reports.

Evaluasi dari uji ulang area lasan yang direpair harus ditabulasikan ke dalam baris baru pada form report.
a.     Jika report asli digunakan , tanda R1, R2, … Rn harus diindikasikan .
b.     Jika form report tambahan digunakan , tanda R harus ditambahkan.

Evaluation of retested repaired weld areas shall be tabulated on a new line on the report form.
a.     If the original report form is used, an R1, R2, … Rn shall prefix the indication number.
b.     If additional report forms are used, the R number shall prefix the report number.




14
Steel Backing.
14
Steel Backing.

UT untuk lasan CJP groove dengan steel backing harus di laksanakan dengan procedure UT yang mengenalkan potensial reflector yang dihasilkan oleh perpotongan antara  base metal dan backing. (lihat Commentary C6.26.12 dari AWS D1.1 untuk  tambahan petunjuk scanning lasan groove yang memakai steel backing).

UT of CJP groove welds with steel backing shall be performed with a UT procedure that recognizes potential reflectors created by the base metal-backing interface (see Commentary C6.26.12 of AWS D 1.1 for additional guidance scanning groove welds containing steel backing).











CALIBRATION REQUIREMENTS LIST
I. EQUIPMENT CALIBRATION AND QUALIFICATION
No.
Description
Reference paragraph
Time  Interval
Remarks
1
Horizontal Linearity
8.1.1
40 hours (one week)

2
Resolution:




Straight Beam
8.2.6
Once (maintain record)


Angle Beam
8.2.4
Once (maintain record)

3
Gain Control
8.1.2
2 months

4
Sound Entry
8.2.1  ~ 8.2.3
8 hours
Angle tolerance
5
Internal reflection
8.1.4
40 hours (one week)






II. CALIBRATION FOR TESTING
1
Horizontal Sweep:




Straight Beam
8.4
Re-calibration see 8.3.4


Angle Beam
8.5
Re-calibration see 8.3.4

2
Angle Beam Zero Reference Level
8.5.2
Re-calibration see 8.3.4

Tidak ada komentar:

Posting Komentar

contoh prosedur asme ultrasonic testing

Daftar Isi   /   Table of Content Par. Uraian Par. Description Page ...